This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of warfarin, dextromethorphan, triprolidine, and tetracaine using XBridge Shield RP18 columns.
Warfarin, an anti-coagulants old as coumadin, is used to inhibit the synthesis of clotting factors. Dextromethorphan is an antitussive drug found in many common cold and cough formulations. Triprolidine is an antihistamine that is found in common allergy formulations. Tetracaine is a local anesthetic of the ester-linkage type, related to procaine.
Columns: |
XBridge Shield RP18 4.6 X 50 mm, 3.5 μm p/n: 186003042 |
Mobile phase A: |
25 mM KH2PO4, pH 7 |
Mobile phase B: |
MeOH |
Flow rate: |
1.0 mL/min |
Injection volume: |
10 μL |
Sample Concentration and Diluent: |
10 μg/mL in H2O |
Temp.: |
30 °C |
Sampling Rate: |
5 points /second |
Detection: |
UV @ 280 nm |
Time Constant: |
1.0 |
Needle Wash: |
5/95 MeOH/H2O |
Instrument: |
Alliance 2695 with 2996 PDA |
Time(min) |
%A |
%B |
---|---|---|
0.0 |
60 |
40 |
7.0 |
20 |
80 |
7.5 |
60 |
40 |
10.0 |
60 |
40 |
WA60203, June 2007