Estrogens are routinely used either as contraceptive medicines or in hormone replacement therapy and they can enter aquatic environments via the discharge of final effluent waters. Estrogens are believed to have a negative effect on aquatic environments by disrupting the hormonal systems of fish. For the European Union, EU Directive 2013/39/EU includes 15 additional priority substances to the water framework Directive 2000/60/EC. In this update, 17α-ethinylestradiol and 17β-estradiol were not included in the priority substance list, but instead added to a watch list in order to gather further data regarding the presence of these compounds in aquatic environments, and the risks they may pose.
This application note summarizes a method for the analysis of estrone, 17α-ethinylestradiol and 17β-estradiol in surface and final effluent waters. To achieve the challenging EU LLOQ parts per quadrillion (ppq) levels required for these compounds, a combination of off-line solid phase extraction (SPE) clean-up and pre-concentration, combined with a large volume injection and triple quadrupole mass spectrometry were utilized.
Application benefits include:
Estrogens are routinely used either as contraceptive medicines or in hormone replacement therapy and they can enter aquatic environments via the discharge of final effluent waters.1 Estrogens are believed to have a negative effect on aquatic environments by disrupting the hormonal systems of fish.1 For the European Union, EU Directive 2013/39/EU2 includes 15 additional priority substances to the water framework Directive 2000/60/EC.3 In this update, 17α-ethinylestradiol and 17β-estradiol were not included in the priority substance list, but instead added to a watch list2 in order to gather further data regarding the presence of these compounds in aquatic environments, and the risks they may pose.
This application note summarizes a method for the analysis of estrone, 17α-ethinylestradiol and 17β-estradiol in surface and final effluent waters. To achieve the challenging EU LLOQ parts per quadrillion (ppq) levels4 required for these compounds, a combination of off-line solid phase extraction (SPE) clean-up and pre-concentration, combined with a large volume injection and triple quadrupole mass spectrometry were utilized.
Extracted samples were prepared and supplied by Scottish Water. Spiked surface water and final effluent samples were first filtered, extracted, and concentrated using an off-line solid phase extraction (SPE) method5-6 (Figure 1). After evaporation and reconstitution in LCMS grade water, the samples were then analyzed by UPLC-MS/MS using a large volume injection (100 µL).
LC system: |
ACQUITY UPLC H-Class with extension loop, needle, and syringe for large volume injection |
Vials: |
TruView LCMS Certified |
Column: |
ACQUITY UPLC BEH C18 1.7 μm, 3.0 x 100 mm |
Column temp.: |
30 °C |
Sample temp.: |
10 °C |
Injection volume: |
100 μL |
Flow rate: |
0.6 mL/min |
Mobile phase A: |
LCMS grade water with 0.01 mM NH4F (analytical grade) |
Mobile phase B: |
50:50 LCMS grade acetonitrile:methanol with 0.01 mM NH4F (analytical grade) |
Time (min) |
Flow rate (mL/min) |
%A |
%B |
---|---|---|---|
Initial |
0.6 |
70 |
30 |
1.00 |
0.6 |
70 |
30 |
3.50 |
0.6 |
5 |
95 |
5.50 |
0.6 |
5 |
95 |
5.60 |
0.6 |
70 |
30 |
8.60 |
0.6 |
70 |
30 |
MS system: |
Xevo TQ-XS |
Ionization mode: |
ESI |
Acquisition mode: |
MRM |
Capillary voltage: |
2.00 kV |
Cone gas flow: |
150 L/Hr |
Desolvation temp.: |
600 °C |
Desolvation gas flow: |
1200 L/Hr |
Nebulizer: |
7 bar |
Data management |
MassLynx v4.2 |
Method optimization was accomplished by evaluating various columns, mobile phase compositions, gradients, and MS transitions. The conditions detailed in the Experimental section provided the best overall performance of those tested. Baseline separation of the target analytes was achieved, and a example of the chromatography from a 50 ng/L solvent standard is shown in Figure 2. Due to the hydrophobic nature of the analytes, they elute in the high organic part of the gradient. As a result, it is challenging to separate the analytes away from matrix components in the final effluent used for this analysis, as shown in the RADAR scan (full scan m/z 100 to 1000) in Figure 3. To minimize any contamination of the MS source, the integrated fluidics system on the Xevo TQ-XS was used to divert the chromatographic region to the MS system for analysis, and unwanted regions to waste.
Satisfactory linearity was achieved for all compounds in matrix matched (spiked post extraction) bracketed calibration curves, over the appropriate ranges. In surface water, a range of 10 to 320 ng/L for 17α-ethinylestradiol, and 62.5 to 2000 ng/L for 17β-estradiol and Estrone provided good linearity (R² >0.998, residuals <15%). An example of the calibration and associated residuals is shown for all three compounds in Figure 4. For final effluent, the matrix matched calibration curves ranged from 120 to 2000 ng/L for all three compounds, also giving acceptable linearity (R²>0.997, residuals <10%). The robustness of the method was assessed using spiked water samples (n=8 for each matrix type) where %RSD values below 6% were obtained.
To evaluate the method performance in surface water, matrix samples were pre-spiked at suitable pg/L (ppq) levels prior to extraction, and prepared in accordance with the methodology detailed in Figure 1. An example of the chromatography and sensitivity observed for a sample pre-spiked at low ppq level in surface water, before clean-up and concentration, is shown in Figure 5.
Matrix effects were determined by quantifying post spiked surface water samples against a solvent calibration curve. For 17β-estradiol and estrone, matrix effects were calculated at ≤-22% (suppression), 17α-ethinylestradiol ≤16% (enhancement). Final effluent, even after the SPE clean-up remained a complex sample, as shown in the RADAR scan in Figure 3, this resulted in significant matrix suppression ≤-72% for all compounds. However even with this significant suppression, low levels of each compound are still detectable.
The recovery and repeatability data for the analytes, spiked at two levels in triplicate in surface water are detailed in Table 2.
The method showed high sensitivity, achieving the target European (2015/495/EU)4 LLOQ (PtP s/n=10) levels for each compound in the surface water matrix. Low level concentrations of all compounds were detected in the final effluent matrix. A standard addition method was used to quantify the analytes present, where 17α-ethinylestradiol was measured at 16.9 pg/L, as shown in Figure 6.
This method highlights the analysis of low ppq levels of synthetic estrogens in surface and final effluent waters. Utilizing Oasis SPE and Sep-Pak SPE sample preparation technologies, the optimized extraction method was found to provide the required concentration and clean-up, giving acceptable recoveries and repeatability in spiked surface water samples. The use of a large volume injection in combination with the ACQUITY UPLC H-Class and Xevo TQ-XS, allowed for the challenging detection requirements of this analysis to be achieved in surface water. The use of standard addition allowed for accurate quantitation of trace residues in final effluent samples.
720006030, Revised November 2020