• Application Note

Qualitative and Quantitative Analysis of Beta-Carotene Using UPC2

Qualitative and Quantitative Analysis of Beta-Carotene Using UPC2

  • Jacquelyn Runco
  • Lakshmi Subbarao
  • Rui Chen
  • Waters Corporation

Abstract

A UPC2 method was successfully developed to separate the three most common carotenoids in less than two minutes. This method is four times faster that traditional methods of analysis, thereby reducing organic solvent consumption by 85%. The high-throughput UPC2 method is ideally suited for laboratories routinely performing quality control and regulatory compliance monitoring where a large number of assays are required.

Benefits

  • A fast UPC2 method to separate the three most common carotenoids that minimizes the risk of degradation. 
  • The UPC2 method is four times faster than the traditional methods of analysis, thereby, reducing organic solvent consumption by 85%.
  • For the trageted analysis, the β-carotene extract in MTBE can be directly injected onto an ACQUITY UPC2 System for analysis without the need for time-consuming evaporation and reconstitution steps.

Introduction

Carotenoids are natural pigments synthesized by plants and some microorganisms. For animals and humans, carotenoids play an important role in vision. Carotenoids also act as important antioxidants with a preventative effect for various diseases.1-2 Since carotenoids cannot be synthesized de novo in the human body, humans need to acquire them through diet and supplements. In 2010, the market value of commercially used carotenoids was estimated to be $1.2 billion and projected to grow to $1.4 billion by 2018.3 As more stringent legislation for regulatory compliance of micronutrients in fortified food products and dietary supplements is being enacted or contemplated, there is an increasing demand for rapid and reliable analytical methods for the analysis and quantification of carotenoids in a variety of matrices.4 The speed of analysis is of particular importance because regulatory compliance monitoring often requires a large number of assays. In addition, many carotenoids are thermal- or photo-sensitive and highly susceptible to isomerization and chemical degradation. Prolonged analysis time could lead to inaccurate quantification results.

Figure 1. Chemical structures and LogP values of three carotenoids used in this study.

Experimental

UPC2 conditions

System:

ACQUITY UPC2

Detection:

PDA detector

Flow rate:

1.5 mL/min

Mobile phase A:

CO2

Mobile phase B:

Ethanol

Column:

Viridis BEH, CSH Fluoro-Phenyl, BEH 2-EP (1.7 μm, 3.0 x 100 mm), and HSS C18 SB (1.8 μm, 3.0 x 100 mm)

Back pressure:

2190 psi

SM temp.:

5 °C

Temp.:

40 °C

Sample diluent:

MTBE

Injection volume:

1 μL

Vials:

Waters Amber Glass 12 x 32 mm Screw Neck Vial, 2 mL

PDA scan range:

220 to 600 nm

Data management:

MassLynx Software

Gradient

Time (min)

B%

0.0

5.0

5.0

20.0

7.0

20.0

8.0

5.0

10.0

5.0

The central part of the carotenoid structure is the long polyene chain of alternating double and single bonds, as shown in Figure 1. Consequently, the carotenoids often possess high hydrophobicity, especially those that do not contain any hetero-atoms, such as lycopene and β-carotene. High-performance liquid chromatography (HPLC) with various absorbance detectors is the most commonly used analytical technique for determining carotenoids qualitatively and quantitatively.1-2, 6-10 Due to their high hydrophobicity, separation of carotenoids by RPLC often results in lengthy analysis times. Furthermore, all RPLC-based methodologies generally suffer from the low solubility of carotenoids in the mobile phase. Non-aqueous reversed phase (NARP) LC has been employed to reduce the run time by using semi-aqueous or non-aqueous mobile phases. However, the NARP approach often involves the use of complex mixtures of organic solvents as the mobile phase. For example, in the official AOAC method for β-carotene in supplements and raw material,11 a mixture of butylated hydroxytoluene (BHT), isopropanol, N-ethyldiisopropylamine, ammonium acetate, acetonitrile, and methanol is used as the mobile phase.

The separation of carotenoids has long been the subject of supercritical fluid chromatography (SFC)12-18 studies since its inception.12 The primary component of the mobile phase in SFC, CO2, offers superior solubility for carotenoids and promotes non-polar interactions between carotenoids and the mobile phase, thereby reducing the retention time.17 In addition to high chromatographic efficiency rendered by the high diffusivity of CO2, the mild temperatures used in SFC are advantageous by avoiding thermal degradation of carotenoids.

UltraPerformance Convergence Chromatography (UPC2) is a new category of separation science that marries the merits of both SFC and UPLC. While adhering to the basic principles of SFC, UPC2 leverages the reduced system volume of UPLC, and more importantly, the exceptional separation power of sub-2-μm particle packed columns, thereby, resulting in a greatly reduced run time, improved resolution, and increased detection sensitivity.

In this application note, we describe fast separations of three common carotenoids by UPC2 in less than 2 minutes. A quantitative analysis of β-carotene dietary supplement capsules is also demonstrated.

Optimized UPC2 conditions for β-carotene extract analysis

Flow rate:

1.5 mL/min

Mobile phase:

75:25 CO2/ethanol, isocratic

Column:

Viridis HSS C18, 1.8 μm, SB 3.0 x 100 mm

Back pressure:

2190 psi

SM temp.:

5 °C

Temp.:

40 °C

Sample diluent:

MTBE

Injection volume:

1 μL

Vials:

Waters Amber Glass 12 x 32 mm Screw Neck Vial, 2 mL

PDA scan range:

350 to 600 nm

Wavelength compensation:

440 nm with a reference wavelength 550 to 600 nm

Sample description

All sample preparation was performed in an environment with subdued lighting. For the column screening and subsequent optimization, 1 mg each of lycopene, β-carotene, and lutein was dissolved in 10 mL of methyl tert-butyl ether (MTBE) to make a 0.1 mg/mL (each) stock solution.

Calibration curve: A serial dilution of a stock solution of β-carotene (0.1 mg/mL in MTBE) was performed. The average peak area of three replicate injections at each concentration was used for each data point. 

Capsule analysis: Three β-carotene capsules with a label claim of 15 mg/capsule were prepared by cutting them open and dissolving the contents in 250 mL of MTBE with slight perturbation. For each assay, six replicate injections were performed, and the average peak area was used for calculating β-carotene content in the capsules.

Results and Discussion

Lutein, lycopene, and β-carotene are the three most common carotenoids found in the North American diet. Preliminary screening work revealed that methanol as mobile phase B (co-solvent) resulted in poor peak shape due to the low solubility of carotenoids in methanol, while isopropanol as a co-solvent led to broader peaks. Ethanol was, therefore, chosen as the co-solvent in all experiments. Figure 2 shows the UPC2/UV chromatograms of the carotenoids mixture from the column screening. The peak identities were confirmed by injecting individual standard using the same condition. While the Viridis C18 SB Column yielded baseline resolution of all three carotenoids, another relatively non-polar column, CSH Fluoro-Phenyl, also provided partial separation between lycopene and β-carotene. No separation between lycopene and β-carotene was observed with either BEH or BEH 2-Ethyl Pyridine columns. Despite the similarities in structure and polarity between lycopene and β-carotene, the octadecyl carbon chains on the ACQUITY UPC2 C18 SB stationary phase offered sufficient resolution to differentiate the two analytes in UPC2.

Figure 2. UPC2/UV chromatograms of a mixture of lycopene, β-carotene, and lutein using the following different columns: (A) HSS C18 SB, (B) CSH Fluoro-Phenyl, (C) BEH 2-EP, and (D) BEH. The identities of the peaks are as follows: 1. Lycopene, 2. β-carotene, and 3. Lutein.
Figure 3. UPC2/UV chromatograms obtained using an Viridis HSS C18 SB Column under different gradient/isocratic conditions including: (A) the initial screening condition: 5% to 20% in 5 min, (B) 20% to 30% in 0.5 min, and (C) 25% isocratic. The identities of the peaks are: 1. Lycopene, 2. β-carotene, and 3. Lutein.

Next, an optimization step was performed to shorten the run time. A ballistic gradient of 20% B/min, shown in Figure 3B, and an isocratic method at 25% B, as seen in Figure 3C, both offered sufficient resolution for all three carotenoids with a run time of less than 2 min. The late-eluting peak (lutein) from the isocratic method has a slightly wider peak width than that from the gradient method, but the isocratic method generated a smoother baseline that can be beneficial for low level detection. The isocratic  method was, therefore, chosen for ensuing quantitative analyses. The optimized method is four times faster than traditional methods of analysis.5 As a result, the organic solvent consumption was reduced by ~85%. It is also important to note that for SFC using C18 columns, retention of non-polar analytes, such as carotenoids, decreases with the analytes’ solubility in the mobile phase.17 Since compressed CO2 offers superior solubility for non-polar analytes, UPC2 is inherently more compatible and faster for carotenoids analyses than RPLC.

Figure 4. UPC2 chromatograms of (A) β-carotene standard, (B) β-carotene extract from capsules, and (C) three carotenoids mixture under optimal chromatographic conditions.

For quantification, the β-carotene content of a commercially available capsule formulation was simply dissolved in MTBE, and the resulting extract was directly injected onto an ACQUITY UPC2 System for analysis using the optimized method, shown in Figure 3C. A representative chromatogram of the resulting β-carotene extract is shown in Figure 4B. T he simple sample preparation exemplifies another advantage of using UPC2 for low polarity sample analysis. Dissolving low polarity samples often requires the use of low polarity solvents, such as MTBE and hexane, which are inherently compatible with UPC2. In contrast, RPLC requires that samples dissolved in low polarity organic solvents be evaporated and reconstituted into suitable diluents prior to analysis.

Figure 5 shows a calibration curve for β-carotene in MTBE with concentrations ranging from 0.0001 to 0.1 mg/mL. The linearity range spans three orders of magnitude with R2 > 0.99. T he limit of detection (LOD, defined as S/N >3) and the limit of quantitation (LOQ, defined as S/N >10) are 50 and 100 ng/mL, respectively. These values are equivalent or better than those reported using HPLC.6-8 The high detection sensitivity can be attributed to, in part, the inherent compatibility between carotene analysis and UPC2. The non-polar interaction between CO2 and β-carotene greatly reduces its retention, thus results in an early eluting sharp peak for improved detection sensitivity.

Figure 5. Calibration curve for β-carotene in MTBE with a concentration range of 0.0001 to 0.1 mg/mL in MTBE.

Tables 1 and 2 summarize the β-carotene capsule analyses. Excellent inter- and intra-assay reproducibility in both retention time and peak area was achieved. Overall assays also yielded good accuracy against the label claim. The sample preparation was simple and straightforward, and the chromatographic analyses using UPC2 were fast and reproducible.

Table 1. Reproducibility of a β-carotene capsule assay with six replicate injections.
Table 2. Quantification of β-carotene in three capsules.

Conclusion

In summary, a UPC2 method was successfully developed to separate the three most common carotenoids in less than two minutes. The method is four times faster than traditional methods of analysis, thereby reducing organic solvent consumption by 85%. The short analysis time also minimizes the risk of on-column degradation of the analytes. The improved speed of analysis is attributed to the inherent compatibility between UPC2 and low polarity analytes. The UPC2 method uses ethanol as the co-solvent instead of mixtures of organic solvents often used in HPLC methods. Thus, the UPC2 method is a much more environmentally sustainable method.

A targeted 1.5-minute UPC2 method was developed for the quantitative analysis of β-carotene in dietary supplement capsules. The dynamic range spans three orders of magnitude, with an LOD and an LOQ of 50 ng/mL and 100 ng/mL, respectively. Using MTBE as the extraction solvent, the resulting β-carotene extract can be directly injected onto an ACQUITY UPC2 System for analysis without the need for time-consuming evaporation and reconstitution steps often associated with RPLC-based methodology. Excellent reproducibility and accuracy were also demonstrated for dietary supplement capsule analysis. The highthroughput UPC2 method is ideally suited for laboratories routinely performing quality control and regulatory compliance monitoring where a large number of assays are required.

References

  1. Rivera SM, Canela-Garayoa R. Analytical tools for the analysis of carotenoids in diverse materials J. Chromatogr. A. 2012; 1224:1-10.
  2. Plozza T, Trenerry VC, Caridi D. The simultaneous determination of vitamins A, E and β-carotene in bovine milk by high performance liquid chromatography– ion trap mass spectrometry (HPLC–MSn). Food Chem. 2012; 134:559-563.
  3. http://www.bccresearch.com/report/carotenoids-global-market-fod025d.html
  4. Blake CJ. Status of Methodology for the Determination of Fat-Soluble Vitamins in Foods, Dietary Supplements, and Vitamin Premixes. J. AOAC Inter. 2007; 90(4):897-910.
  5. Hung PV, Hatcher DW. Ultra-performance liquid chromatography (UPLC) quantification of carotenoids in durum wheat: Influence of genotype and environment in relation to the colour of yellow alkaline noodles (YAN). Food Chem. 2011; 125:1510-1516.
  6. Mitrowska K, Vincent U, von Holst C. Separation and quantification of 15 carotenoids by reversed phase high performance liquid chromatographic coupled to diode array detection with isobestic wavelength approach. J. Chromatogr. A. 2012; 1233:44-53.
  7. Chauveau-Duriot B, Doreau M, Noziere P, Graulet B. Simultaneous quantification of carotenoids, retinol, and tocopherols in forages, bovine plasma, and milk: validation of a novel UPLC method. Anal. Bioanal. Chem. 2010;397: 777.
  8. Santos J, Mendiola JA, Oliveira MBPP, Ibanez E, Herrero M. Sequential determination of fat- and water-soluble vitamins in green leafy vegetables during storage. J. Chromatogr. A. 2012; 1261:179-188.
  9. Epriliati I, Kerven G, D’Arcy B, Gidley MJ. Chromatographic analysis of diverse fruit components using HPLC and UPLC. Anal. Methods. 2010; 2:1606-1613.
  10. Granado-Lorencio F, Herrero-Barbudo C, Blanco-Navarro I, Perez-Sacristan B. Suitability of ultra-high performance liquid chromatography for the determination of fat-soluble nutritional status (vitamins A, E, D, and individual carotenoids) Anal. Bioanal. Chem. 2010; 397:1389-1393.
  11. AOAC official method 2005.07, β-Carotene in Supplements and Raw Materials Reversed-Phase High Pressure Liquid Chromatographic Method, First Action 2005.
  12. Giddings JC, McLaren L, Myers MN. Dense-gas chromatography of nonvolatile substances of high molecular weight. Science. 1968; 159:197-199.
  13. Aubert M-C, Lee CR, Krstulovic AM. Separation of trans/cis- α- and β-carotenes by supercritical fluid chromatography. (I) Effects of temperature, pressure and organic modifiers on the retention of carotenes. Journal of Chromatogr. 1991; 557:47-58.
  14. Lesellier E, Tchapla A, PeHchard M-R, Lee CR, Krstulovic AM. Separation of trans/cis α- and β-carotenes by supercritical fluid chromatography. (II) Effect of the type of octadecyl-bonded stationary phase on retention and selectivity of carotenes. Journal of Chromatogr.1991; 557:59-67.
  15. Lesellier E, Tchapla A, Marty C, Lebert A. Analysis of carotenoids by high-performance liquid chromatography and supercritical fluid chromatography. Journal of Chromatography. 1993; 633:9-23.
  16. O’Neil C and Schwartz SJ. Chromatographic analysis of cis/trans carotenoid isomers. Journal of Chromatography. 1992; 624:235-252.
  17. Sakaki K, Shinbo T, Kawamura M. Retention behavior of β-carotene on polar and non-polar stationary phases in supercritical fluid chromatography. Journal of Chromatographic Science. 1994; 32:172-177.
  18. Tee E-S and Lim C-L. The analysis of carotenoids and retinoids: a review. Food Chemistry. 1991; 41:147-193.

720004550, January 2013