For research use only. Not for use in diagnostic procedures.
This application note demonstrates improvements to the metabolic profiling workflow by coupling RAMMP with a novel DIA method (SONAR), providing highly specific and unbiased two-dimensional metabolomic data.
Previous studies utilizing rapid microbore metabolic profiling (RAMMP) have shown comparable group discrimination and improved selectivity over conventional UPLC chromatography.1 Here, we demonstrate further improvements to the workflow by coupling RAMMP with a novel DIA method (SONAR), providing highly specific and unbiased two-dimensional metabolomic data. SONAR is an acquisition technique comprised of a low-resolution quadrupole mass filter, which is scanned repetitively and both precursor and MS-MS data are acquired at spectral rates approaching 2000 spectra/s. Sample sets consisting of urine collected from pregnant women over three trimesters were used to demonstrate SONAR for use with high throughput analyses. Data were analyzed and interrogated using Progenesis QI, while targeted quantitation was provided using Skyline.
Urine samples (Innovative Research Inc.) were prepared as previously described.2 Briefly, particulates and debris were removed by centrifuging at 10,000 g for 10 minutes prior to diluting 2-fold with water. Samples were vortexed and transferred to glass vials in preparation for LC-MS analysis.
LC system: |
ACQUITY UPLC M-Class |
Column: |
BEH C18 1.7 μm, 1.0 mm x 100 mm or 300 μm x 100 mm Reversed Phase Analytical |
Column temp.: |
50 °C |
Flow rate: |
100 μL/min (1 mm I.D.)/7 μL/min (300 μm I.D.) |
Mobile phase: |
(A) Water/0.1% formic acid; (B) Acetonitrile/0.1% formic acid |
Gradient: |
1% to 95% B over 12, 6, or 3 min |
Injection volume: |
5 μL |
MS system: |
Xevo G2-XS |
Ionization mode: |
ESI (+) at 2.2 kV; ESI (-) at 1.7 kV |
Cone voltage: |
30 V |
Acquisition mode: |
SONAR 50 m/z to 1200 m/z both functions (low and elevated energy) |
Quadrupole settings: |
12 Da window operating over 250–800 Da |
Acquisition rate: |
Low and elevated energy functions at 0.1 s |
Collision energy: |
Low energy function at 5 eV and elevated energy function from 20 eV to 50 eV (positive ion) and 25 eV to 55 eV (negative ion) |
Resolution: |
30,000 FWHM |
Progenesis QI |
EZInfo |
MassLynx |
Skyline |
Spotfire |
The LC-MS metabolite data were processed and searched with Progenesis QI. Normalized label-free quantification was achieved with additional statistical analysis conducted using EZInfo (Umetrics, Sweden). Compound searches were conducted using METLIN and HMDB. Quantitative analysis was performed with Skyline (University of Washington) using libraries derived from Progenesis QI compound searches.
SONAR DIA acquisition provides multi-dimensional data sets, exhibiting improved specificity. Figure 1 represents typical SONAR data and demonstrates that the format is the same as other multidimensional data sets, e.g. ion mobility; hence, exhibits improved specificity. Urine based data were acquired using either a 1 mm or a 300 µm I.D. column. Figure 2 provides example chromatographic urine profiles generated using RAMMP (3 min gradient) and conventional (12 min gradient) methods.
An evaluation of the number of peak-detected features achieved for the different column configurations, gradient lengths, and quadrupole window was conducted (Figure 3). An increase of approximately 50% is observed with decreasing column diameter, when comparing against the same gradient and quadrupole window. Unsupervised principal component analysis (PCA) highlights differentiation of the three trimesters regardless of the gradient selected (Figure 4). To ensure robustness and consistency of the results when switching between conventional and RAMMP based methods, the discriminating features responsible for the PCA based separation were assessed for both scenarios (Figure 5).
High specificity provided by SONAR reduces potential interference effects and thereby increases quantitative confidence. A number of metabolites based on the RAMMP method (3 min) were selected for targeted analysis using open source Skyline informatics (Figure 6). Precursor/product ions list were provided to the software, along with quadrupole (precursor) m/z extraction information.
Achieving an adequate number of scans is imperative for maintaining quantitative precision. Figure 7 demonstrates the fast scanning capabilities of SONAR for both precursor and product ions. This example shows urothion being acquired over various scan rates (0.5, 0.3, and 0.1 sec). Applying a 0.1 sec scan provides more than 10 points over the chromatographic peak (1.7 sec FWHM) and generates the expected 2.5-fold ratio when comparing transitions for first and third trimester cohorts from Figure 6. Comparing three representative transitions shows consistency with scan rate providing additional confidence and increased quantitative precision.
720006072, August 2017