In this application note, USP methods from different lidocaine formulations were modernized by switching from older Atlantis dC18 and XTerra C18 Columns to contemporary XBridge BEH C18 Columns. Improved method speed and resolution originated from both the change in particle and the change in particle size.
The analysis of pharmaceutical semisolid dosage forms, such as jellies, ointments, and creams, can be difficult due to excipients used in their formulation. In contrast, oral and injectable solutions are simpler and therefore easier to analyze. Lidocaine, an anesthetic used to treat pain from certain procedures, comes in both semisolid and solution forms. In this application note, we discuss the modernization of the USP liquid chromatography (LC) methods, using XBridge Columns, for both types of lidocaine formulations. In this context, modernization is chromatographic analysis speed improvement through the use of smaller particles, shorter columns, and higher flow rates, without sacrificing performance. Herein we also demonstrate equivalent outcomes across various LC instruments from different manufacturers.
Lidocaine hydrochloride jelly and lidocaine ointment
Diluent: 0.1% aq phosphoric acid/acetonitrile (50:50, v/v)
System suitability solution (USP assay method): 0.1 mg/mL of lidocaine RS (Reference Standard), 1, and 0.04 mg/mL of ropivacaine related compound A RS, 2, in diluent.
Sample solution for lidocaine hydrochloride jelly (USP assay method): Generic lidocaine hydrochloride jelly (2%, w/v) was used in this study. The sample solution was prepared as per the lidocaine hydrochloride jelly USP method. Specifically, 0.3 mL of jelly was transferred to a 50 mL volumetric flask containing about 25 mL of diluent. The solution was sonicated for 5 minutes. Further diluent was added to the volume mark to obtain a lidocaine hydrochloride concentration of 0.12 mg/mL.
Sample solution for lidocaine ointment (USP assay method): Generic lidocaine ointment (5%, w/w) was used in this study. The sample solution was prepared as per the lidocaine ointment USP method. Specifically, 100 mg of ointment was weighed into a 50 mL volumetric flask. Diluent (25 mL) was added and the mixture was sonicated for 10 minutes. Further diluent was added to the volume mark to obtain a lidocaine concentration of 0.1 mg/mL.
Lidocaine hydrochloride oral topical solution
Buffer: aq monobasic potassium phosphate buffer, 4.85 g/L, adjusted to pH 8.00 with 10 N sodium hydroxide.
Diluent: buffer/acetonitrile (70:30, v/v).
System Suitability Solution (USP assay method): 0.0043 mg/mL of lidocaine RS, 1, 0.00065 mg/mL of ropivacaine related compound A RS, 2, and 0.005 mg/mL of lidocaine related compound H, 3, in diluent.
Sample solution for lidocaine hydrochloride oral topical solution (USP assay method): Generic lidocaine hydrochloride oral topical solution (2%) was used in this study and was prepared as per the USP lidocaine hydrochloride oral topical solution method. Specifically, 2.0 mL of topical solution was added to a 20 mL vial containing 6 mL of mobile phase and mixed well to obtain a concentration of 5 mg/mL of lidocaine hydrochloride. A 1.0 mL aliquot of this solution was transferred to another 20 mL vial containing 4 mL of mobile phase to get a concentration of 1 mg/mL of lidocaine hydrochloride.
Instruments: |
Alliance e2695 Quaternary HPLC with 2489 UV/Visible Detector |
|
Shimadzu Nexera-I Quaternary LC 2040C 3D |
|
Agilent 1100 Binary LC with Agilent 1100 DAD Detector |
|
Agilent 1260 Infinity Quaternary LC with Agilent 1260 DAD Detector |
Data management: |
Empower 3 CDS Software |
Columns: |
Atlantis dC18, 5 μm, 4.6 x 150 mm (p/n: 186001344) (USP Method, L/dp=30,000) |
|
XBridge BEH C18, 5 μm, 4.6 x 150 mm (p/n: 186003116) (Modernized Method, L/dp = 30,000) |
|
XBridge BEH C18, 3.5 μm, 4.6 x 150 mm (p/n: 186003034) (Modernized Method, L/dp = 42,800) |
|
XBridge BEH C18 XP, 2.5 μm, 4.6 x 75 mm (p/n: 186006038) (Modernized Method, L/dp = 30,000) |
Mobile phase A: |
0.1 % aq phosphoric acid |
Mobile phase B: |
Acetonitrile |
Composition profile: |
10% B to 90% B in 10 min (Atlantis dC18, 5 μm, 4.6 x 150 mm) |
|
10% B to 90% B in 10 min (XBridge BEH C18, 5 μm, 4.6 x 150 mm) |
|
10% B to 90% B in 7 min (XBridge BEH C18, 3.5 μm, 4.6 x 150 mm) |
|
10% B to 90% B in 2.5 min (XBridge BEH C18 XP, 2.5 μm, 4.6 x 75 mm) |
Flow rate: |
0.8 mL/min (Atlantis dC18, 5 μm, 4.6 x 150 mm) |
|
0.8 mL/min (XBridge BEH C18, 5 μm, 4.6 x 150 mm) |
|
1.14 mL/min (XBridge BEH C18, 3.5 μm, 4.6 x 150 mm) |
|
1.6 mL/min (XBridge BEH C18 XP, 2.5 μm, 4.6 x 75 mm) |
Column temp.: |
25 °C |
Detection (UV): |
210 nm |
Injection volume: |
5 μL (Atlantis dC18, 5 μm, 4.6 x 150 mm) |
|
5 μL (XBridge BEH C18, 5 μm, 4.6 x 150 mm) |
|
5 μL (XBridge BEH C18, 3.5 μm, 4.6 x 150 mm) |
|
2.5 μL (XBridge BEH C18 XP, 2.5 μm, 4.6 x 75 mm) |
Columns: |
XTerra Shield RP18, 3.5 μm, 4.6 x 150 mm (p/n: 186000536) (USP Method, L/dp = 42,800) |
|
XBridge BEH Shield RP18, 3.5 μm, 4.6 x 150 mm (p/n: 186003045) (Modernized Method, L/dp = 42,800) |
|
XBridge BEH Shield RP18 XP, 2.5 μm, 4.6 x 100 mm (p/n: 186006063) (Modernized Method, L/dp = 40,000) |
Mobile phase A: |
Buffer |
Mobile phase B: |
Acetonitrile |
Composition profile: |
30% B (XTerra RP18, 3.5 μm, 4.6 x 150 mm) |
|
30% B (XBridge BEH Shield RP18, 3.5 μm, 4.6 x 150 mm) |
|
30% B (XBridge BEH Shield RP18 XP, 2.5 μm, 4.6 x 100 mm) |
Flow rate: |
1.0 mL/min (XTerra RP18, 3.5 μm, 4.6 x 150 mm) |
|
1.0 mL/min (XBridge BEH Shield RP18, 3.5 μm, 4.6 x 150 mm) |
|
1.4 mL/min (XBridge BEH Shield RP18 XP, 2.5 μm, 4.6 x 100 mm) |
Column temp.: |
45 °C |
Detection (UV): |
230 nm |
Injection volume: |
20 μL (XTerra RP18, 3.5 μm, 4.6 x 150 mm) |
|
20 μL (XBridge BEH Shield RP18, 3.5 μm, 4.6 x 150 mm) |
|
13.3 μL (XBridge BEH Shield RP18 XP, 2.5 μm, 4.6 x 100 mm) |
Lidocaine solution1 and semisolid formulations,2,3 sample solutions were analyzed using XTerra C18, 3.5 µm, 4.6 x 150 mm (p/n: 186000536) and Atlantis dC18, 5 µm, 4.6 x 150 mm (p/n: 186001344) compendial columns, respectively, with USP methods as specified in their respective monographs. The Alliance and Agilent 1100 HPLCs plus the Agilent 1260 and Shimadzu Nexera-i UHPLCs were used. As shown in Figure 1 and Figure 2, similar chromatograms were observed across all the instruments, with slight variation in the retention times which may be due to system volume differences.
Even though semisolids are considered to be more difficult formulations to work with, no challenges were encountered during the analysis of these formulations. Hence, method modernization was carried out on an Alliance HPLC System using the system suitability solution. The modernized methods were then demonstrated across instruments from different vendors as discussed below.
We can modernize4 a USP LC method by scaling to smaller particle sizes. At a given column diameter, the speed of an LC method can be expressed as the delivery of the required number of mobile phase column volumes,5 # CVs, per time interval, t, shown in eq 1. Thus method speed increases with flow rate, F, and decreases with column length (via the lower calculated column void volume, V0).
LC method speed = (#CVs)/t = F/V0 - eq. 1
For isocratic USP methods, the USP General Chapter <621> specifies the allowed method modernization changes in particle size, flow rate, and column geometry.6 Eq 2 gives the relationship between the flow rates F1 and F2, the column diameters dc1 and dc2, and the particle sizes dp1 and dp2 for the original and modernized conditions, respectively.
F2 = F1 • [(dc2)2 • dp1/(dc1)2 • dp2] - eq. 2
Decreasing the particle size alone allows a faster flow rate per USP guidelines, assuming the pressure ceiling of the LC instrument is not reached. A smaller particle size also permits use of a shorter column as long as the original ratio of column length, L, to particle size, dp, is maintained in the modernized method. This “equivalent L/dp” USP guideline7 provides an additional method speed increase as the particle size decreases since smaller particles can give equivalent efficiency using higher flow rates and shorter columns.
The USP method for lidocaine ointment and lidocaine hydrochloride jelly specifies an older column, Atlantis dC18, using a gradient method. USP General Chapter <621> guidelines do not allow changes to a gradient chromatographic method without a revalidation. However, if a faster analysis is desired, modernization can be performed and then the resulting method subjected to the USP validation,8,9 process.
We began the method modernization by switching from the Atlantis dC18, 5 µm, 4.6 x 150 mm (p/n: 186001344) compendial column to the more recent L1 (C18), XBridge BEH C18, 5 µm, 4.6 x 150 mm (p/n: 186003116) Column. This is typically done to gain access to more robust column particles such as the BEH hybrid technology. Figure 3a and Figure 3b shows the chromatograms observed for these columns with the original USP method. The XBridge Column gives an 11% shorter analysis time (due to reduced analyte retention) and improved resolution between the two closely eluting peaks from lidocaine, 1, and ropivacaine related compound A, 2 (due to increased selectivity of the BEH particle). The method was then scaled to a smaller particle size, by switching to an XBridge BEH C18, 3.5 µm, 4.6 x 150 mm (p/n: 186003034) Column. There is a change of only the particle size so the method speed increases due to a scaled flow rate adjustment alone, per eq 2. Specifically, the flow rate rises from 0.8 mL/min to 1.14 mL/min, affording a 38% analysis time reduction from the compendial conditions, Figure 3c. The USP method was further modernized to an XBridge BEH C18 XP, 2.5 µm, 4.6 x 75 mm (p/n: 186006038) Column. This is an “equivalent L/dp” type of modernization which produces a larger increase in method speed. The analysis time drops by ca. 73% from the original method without compromising the resolution, as shown in Figure 3d.
The fastest modernized method, using the XBridge BEH C18 XP, 2.5 µm, 4.6 x 75 mm (p/n: 186006038) Column, was run on Agilent 1100, Agilent 1260, and Shimadzu Nexera-i instruments with the semisolid formulations system suitability solution. The results are shown in Figure 4. The Alliance HPLC and Agilent 1260 instruments gave the narrowest peaks whereas the Agilent 1100 and Shimadzu Nexera-i instruments gave wider peaks. The Agilent 1260 and particularly the Shimadzu Nexera-i provided higher backpressures for the same column and method, consistent with these being UHPLC class instruments.
The USP method for the lidocaine solution formulation also recommends an older column, XTerra RP18, 3.5 µm, 4.6 x 150 mm (p/n: 186000536). Although XTerra Columns have more sturdy particles compared to predecessors, we also modernized by transferring this USP method to an even better column particle, BEH. This change resulted in improved resolution between ropivacaine related compound A, 2, and lidocaine related compound H, 3, as depicted in Figure 5a and Figure 5b. An “equivalent L/dp” modernization of the compendial method requires use of a 2.5 µm, 4.6 x 100 mm column. This modification provided the chromatogram in Figure 5c with a 50% decrease in the analysis time.
The system suitability solution was run with the modernized scaled method using the XBridge BEH Shield RP18 XP, 2.5 µm, 4.6 x 100 mm (p/n: 186006063) Column on different LC instruments. Similar results were obtained across these instruments, as shown in Figure 6.
In this application note, USP methods from different lidocaine formulations were modernized by switching from older Atlantis dC18 and XTerra C18 Columns to contemporary XBridge BEH C18 Columns. Improved method speed and resolution originated from both the change in particle and the change in particle size. Running the modernized methods on four LC instruments from three different vendors afforded comparable excellent separations regardless of the system used. This demonstrates the ability to achieve quicker analytical answers and higher sample throughput with such modern robust columns when moving from 5 µm to 3.5 µm to 2.5 µm particles on various manufacturers’ LC instruments.
720006179, January 2018