This application note describes EPA Method 8330.0 determination of nitroaromatics and nitramines by high performance liquid chromatography (HPLC).
The presence of numerous military and defense sites around the world, both active and decommissioned, has resulted in the presence of explosives compounds in locations where they can enter the water supply. In the US, the evaluation of sites for potential contamination is carried out by the United States Environmental Protection Agency (US EPA), US Department of Defense, and US Department of Energy in support of Superfund, RCRA, and Base Closure environmental programs.
Instrument: |
Waters Alliance HPLC system with 2487 Dual λ Absorbance detector |
Eluent: |
10 mM ammonium formate/isopropanol |
Column: |
XTerra Phenyl, 3.5 μm, 2.1 x 150 mm @ 40 °C |
Injection: |
10 μL of standard |
Flow rate: |
0.25 mL/min |
Detection: |
UV @ 254 nm |
Data: |
Waters Empower software |
Sample Matrix: |
Groundwater and surface water, low concentration |
Sample Prep: |
Solid-phase extraction using a 500 mL sample Porapak reverse-phase sorbent (RDX), elute with acetonitrile |
Oasis HLB Extraction Method
Oasis HLB Extraction Cartridge, 6 cc, 200 mg
Dilute 100 μL of AccuStandard mix (M-8330-R) to 10 mL with eluent for a working 10 ppm standard mixture.
Dissolve 0.631 g of ammonium formate in 100 mL water. Transfer to a 1 L volumetric flask. Add 200 mL of Isopropanol. Dilute to the mark with water and mix well. Carefully pH to 3.8 with formic acid, then filter and degas.