In this study, we combine two compendial methods into one UPLC method for the separation of ziprasidone HCl impurities to reduce analysis time and improve chromatographic separation.
The majority of today’s compendial methods are considered outdated by both industry and regulators, and do not take advantage of the recent advances in recent chromatographic instrumentation and stationary phase solutions. The U.S. Pharmacopeia (USP) launched an initiative in May 2010 to modernize monographs for small molecule drug products and excipients, identified as a priority by the U.S. Food and Drug Administration, that use outdated technology, have safety or environmental concerns, or are missing key aspects.
HPLC methods listed in the United States Pharmacopeia (USP) National Formulary (NF) monographs are typically long and consume a high volume of solvents. With the evolution of new technology, many companies want to migrate to UPLC technology with sub-2-µm particle columns to reduce analysis time and solvent usage, improve chromatographic performance, and maintain quality. Adopting UPLC technology streamlines laboratory processes by improving efficiency, productivity, and profitability of pharmaceutical manufacturing facilities.
Ziprasidone HCl is an anti-psychotic drug administered orally to treat acute manic or mixed episodes associated with bipolar disorder. The USP specifies two HPLC methods1 to analyze impurities of ziprasidone HCl. In this study, we combine two compendial methods into one UPLC method for the separation of ziprasidone HCl impurities to reduce analysis time and improve chromatographic separation. Performance of the UPLC method is measured by evaluating five replicate injections of the system suitability solution against the requirements listed in the USP monograph for ziprasidone HCl.
All solutions, shown in Table 1, were prepared in diluent (methanol/water/HCl at 20:5:0.01) as per the impurities methods defined in the USP monograph for ziprasidone HCl.1 Since the USP does not list a sample preparation protocol for the Ziprasidone HCl capsules, the drug product sample preparation was used with one modification. Sample solutions were filtered through 0.2-µm PTFE syringe filters to remove any particulates.
Empower 3 Software
HPLC conditions for late-eluting impurities method |
|
---|---|
LC system: |
Alliance 2695 HPLC with 2489 UV/Visible Detector |
Column: |
XBridge C8 4.6 x 150 mm, 5 μm |
Column temp.: |
35 °C |
Sample temp.: |
10 °C |
Injection volume: |
20 μL |
Flow rate: |
1.0 mL/min |
Mobile phase: |
11:1:8 acetonitrile/methanol/buffer |
Buffer: |
50 mM potassium phosphate monobasic, pH 6.0 adjusted with 5N potassium hydroxide |
Separation mode: |
Isocratic |
Wash solvents: |
50:50 water/methanol |
Detection: |
UV, 229 nm |
For five replicate injections of the system suitability solution:
LC system: |
Alliance 2695 HPLC with 2489 UV/Visible Detector |
Column: |
XBridge C8 4.6 x 150 mm, 5 μm |
Column temp.: |
35 °C |
Sample temp.: |
10 °C |
Injection volume: |
20 μL |
Flow rate: |
1.0 mL/min |
Mobile phase: |
11:1:8 acetonitrile/methanol/buffer |
Buffer: |
50 mM potassium phosphate monobasic, pH 6.0 adjusted with 5N potassium hydroxide |
Separation mode: |
Isocratic |
Wash solvents: |
50:50 water/methanol |
Detection: |
UV, 229 nm |
For five replicate injections of the system suitability solution:
LC system: |
ACQUITY UPLC H-Class with ACQUITY UPLC TUV Detector |
|
Column: |
ACQUITY UPLC BEH C8 2.1 x 50 mm, 1.7 μm |
|
Column temp.: |
35 °C |
|
Sample temp.: |
10 °C |
|
Injection volume: |
1.4 μL |
|
Flow rate: |
0.5 mL/min |
|
Mobile phase: |
Solvent A: 50mM Potassium phosphate monobasic, pH 6.0 |
|
Solvent B: |
Methanol |
|
Separation mode: |
Gradient |
|
Wash solvents: |
50:50 water/methanol |
|
Detection: |
UV, 229 nm |
Step |
Time (minutes) |
%A |
%B |
---|---|---|---|
1 |
Initial |
65.0 |
35.0 |
2 |
0.60 |
35.0 |
65.0 |
3 |
2.70 |
25.0 |
75.0 |
4 |
3.90 |
25.0 |
75.0 |
5 |
3.96 |
65.0 |
35.0 |
6 |
6.50 |
65.0 |
35.0 |
Two compendial methods for the early and late eluting impurities of ziprasidone HCl were analyzed as described using the Alliance 2695 HPLC System equipped with a 2489 UV/Visible Detector. The USP designates an L7 column, specifically a Zorbax RX-C8 column, for the impurities testing. Using the Waters Reversed-Phase Column Selectivity Chart (www.waters.com/selectivitychart), an equivalent Waters XBridge C8 Column was chosen. Chromatographic data for both HPLC methods are displayed in Figure 1.
Method development of a UPLC-based separation of ziprasidone HCl and all the specified related compounds was conducted by a gradient elution using a potassium phosphate monobasic buffer with pH of 6.0 and methanol. The mobile phase with a neutral pH 6.0, specified in the late-eluting impurities method, was investigated due to the neutral characteristics of the ziprasidone HCl related substances.
A Waters ACQUITY UPLC BEH C8 Column has the same stationary phase as the Waters HPLC XBridge C8 Column, hence it was selected for method development. The dimensions of the UPLC Column were decided by the column length (L) to the particle size (dp) ratio (L/dp). The L/dp value for the HPLC column is 30,000. The L/dp value for the UPLC Column with a dimension of 50 mm in length and 1.7-µm particle size is 29,412. Injection volume was scaled down to UPLC using the Waters ACQUITY UPLC Columns Calculator, as previously described.2
A generic scouting gradient from 5% to 95% methanol over 15 minutes was performed to investigate the elution and separation between the peaks. Increasing the starting percent of the organic solvent to 35% lowered the run time, providing an adequate separation between all the peaks. Examples of the chromatographic data for the gradient elution study are displayed in Figure 2.
Different flow rates were also explored to further reduce run time and determine the effect on resolution. The flow rate was increased from 0.3 to 0.5 mL/min, as shown in Figure 3. It was observed that an increase in flow rate lowered the resolution between peaks 2 (related compound B) and 3 (ziprasidone), but still passing the USP criteria for the system suitability, shown in Table 2. In addition, a flow rate of 0.5 mL/min reduced run time from 6.6 to 4.0 minutes and decreased peak width by 32%.
The resulting UPLC method successfully resolved ziprasidone and all the specified related compounds, shown in Figure 4.
Performance of the UPLC method was determined by comparing the system suitability results of the five replicate injections of the system suitability solutions against the acceptance criteria defined in the USP monograph for ziprasidone HCl. System suitability results met the USP requirements for the early and late impurities methods listed in the USP monograph for ziprasidone HCl, shown in Table 2.
Table 2. System suitability results for five replicate injections of the system suitability solution on an ACQUITY UPLC H-Class System.
*Note: Elution order for related compounds C and D has changed with analysis on UPLC. Value reported in the table reflects resolution between ziprasidone and related compound D
720004639, May 2013