This is an Application Brief and does not contain a detailed Experimental section.
Waters presents some critical column and chemistry best practices for performing successful oligonucleotide analysis by ion-pair reversed-phase (IP-RP) liquid chromatography. Ensuring high quality oligonucleotides, either in diagnostic applications or as therapeutic entities, relies on robust analytical methods. For example, in response to the global COVID-19 pandemic, PCR-based diagnostic kits have been developed to detect SARS-CoV-2 genetic code in novel coronavirus patients. Accurate viral detection via PCR requires high quality oligonucleotide primers and probes. Additionally, oligonucleotides are being investigated as therapeutics (including mRNA-based vaccines) for the treatment and prevention of COVID-19.
In this application brief, Waters provides a short list of best practices to characterize oligonucleotides by ion-pair reversed-phase (IP-RP) liquid chromatography. These best practices cover column selection, separation temperature, pH optimization, pore size selection, mobile phase choice, rapid method development considerations, purification guidelines, and popular buffer recipes. Following these best practices for oligonucleotide analysis helps ensure robust methods, enabling the delivery of consistently high quality oligonucleotides for therapeutic or diagnostic applications.
The COVID-19 pandemic serves as an example of the importance of oligonucleotides in diagnostic applications. Timely diagnosis of SARS-CoV-2 viral infection remains paramount to successfully managing the novel coronavirus pandemic. As a result, many firms have developed in vitro diagnostic tests to detect the presence of SARS-CoV-2 genetic material. Of the emergency use authorizations (EUA) for in vitro diagnostic tests in the United States, many of them are PCR-based (e.g. qPCR, RT-PCR) tests which detect SARS-CoV-2 viral genetic code.1 Diagnostic kits indicated for the molecular detection of viral genetic code rely on oligonucleotides as primers and probes during PCR amplification and detection. As a result, analytically assessing the quality and ensuring the appropriate purity of oligonucleotide primers and probes is of principle importance for a successful diagnostic test.
Waters created a short list of best practices that have emerged as a result of reviewing method and product development resources for IP-RP oligonucleotide separations. We discuss 7 key practices below.
1) Consider the use of ACQUITY PREMIER Oligonucleotide BEH C18 Columns with MaxPeak High Performance Surfaces (HPS). PREMIER column hardware built with MaxPeak HPS creates a barrier between the metal surfaces of the fluid path and sample thus minimizing interaction between analytes and column hardware. Employing this technology:
i. Eliminates the need for time consuming, costly, and transient column conditioning with sacrificial oligonucleotide sample.
ii. Improves the quality and repeatability of oligonucleotide analyses for both conventional LC-UV applications and, in particular for, sensitive LC-MS analysis.
For more detailed information, see the Waters Application Note: Improved Chromatographic Analysis of Oligonucleotides with ACQUITY PREMIER Oligonucleotide BEH C18 Columns. 720007012EN.
2) Perform oligonucleotide separations at elevated pH and temperature for best results.
i. Elevated temperature (60 °C) prevents the oligonucleotide secondary structure from impacting retention. For CG rich or G-rich oligonucleotides with high degree of secondary structure, it may be necessary to increase column temperature to 80 or 90 °C.
ii. High pH buffers (pH ≥7) are commonly used in oligonucleotide separations (e.g. TEAA).
iii. TEA-HFIP was found to be a robust mobile phase offering superior LC-MS sensitivity and resolution across various sized single stranded oligonucleotides.2
iv. The high pH used for IP-RP oligonucleotide separations renders most common silica-based stationary phases unsuitable.
v. Waters BEH sorbent technology lends itself well for oligonucleotide separation due to its high pH stability and temperature tolerance.
For more detailed information, see Waters HPLC and UPLC Columns for the Analysis of Oligonucleotides. 720002376EN.
3) Choose a relevant pore size for your oligonucleotide separation. Choosing the correct pore size enables appropriate analyte diffusivity resulting in the best interaction between the oligonucleotide and the ligand. Improved ligand interactions improve peak shape.
i. 130 Å pore size is ideally suited for single stranded oligonucleotides (2–100 mers).
ii. 300 Å pore size allows for efficient separation of both single stranded oligonucleotides and longer dsDNA fragments. Even though labeled as a peptide column, scientists have been successful with an ACQUITY PREMIER Peptide BEH C18, 300 Å Column (186009495, 186009494, 186009493).
4) Choose an appropriate mobile phase. See section 7 for buffer recipes.
i. Triethylamine/hexafluoroisopropanol (TEA/HFIP) is MS compatible and has impressive resolving power. Higher TEA/HFIP buffer concentrations improve separation performance. Lower concentrations improve MS sensitivity.
ii. Hexylammonium acetate (HAA) also offers exceptional resolution and MS compatibility. However, the MS compatibility of HAA is less than that of TEA/HFIP. Use of HAA may result in better separation of labeled oligonucleotides and longer oligonucleotides (>35-mer) compared to TEA/HFIP. This may be relevant if performing only LC analysis.
iii. Fresh TEA/HFIP and HAA/HFIP mobile phases are critical to good separations. These semi-volatile mobile phases can gradually lose their separation strength and MS spectra become contaminated with alkali ion adducts. For robust day-to-day results, make mobile phases daily or in limited quantities. Upper limit of mobile phase usability is one week.
iv. Both TEA and HFIP should be prepared in a fume hood, use Waters APC solvent bottle caps to prevent gassing out, and if possible, use a snorkel above the system.
For more detailed information, see the below Waters Application Notes
5) Quick three step method development for <5-minute IP-RP oligonucleotide separations.
i. Identify suitable initial gradient strength or start with a scouting gradient. With 15 mM TEA/400 mM HFIP ion pairing system, an example scouting gradient may be:
ii. Adjust gradient slope to achieve desired separation (shallower gradients increase resolution, but increase time needed for analysis). Adjust the starting percentage of MeOH to reduce the time of analysis as needed. Extend the MeOH gradient time as needed until the target oligonucleotide and impurities are eluted. High organic flush is then inserted to elute highly retained components (often non-oligonucleotide components) and minimize carryover. Target oligonucleotides should elute during the gradient and not in the high organic flush.
iii. If speed is important, speed up the separation by increasing the flow rate while proportionally reducing gradient time (constant gradient column volume). Selectivity of separation should not change while minimal loss in resolution can be observed. When driving towards sensitivity or a need for optimal resolution, use lower flow rates.
Figure 4. Separation of 30 to
60 nt oligodeoxythymidines using 2.1 x 50 mm, 1.7 μm ACQUITY UPLC OST C18
Column.
For more detailed information, see the below Waters Application Notes:
6) Guidelines for oligonucleotide purification using XBridge Oligonucleotide BEH C18 Columns.
i. XBridge Oligonucleotide BEH C18, 130 Å Columns are the preferred offering for detritylated oligonucleotide purifications due to the availability of column sizes designed to meet lab-scale isolation requirements.
ii. The choice of XBridge Oligonucleotide C18 Column dimension and operating flow rate depends primarily on the scale of the synthesis reaction mixture.
*Values are approximate and
vary depending on oligonucleotide length, base composition, and “heart-cutting”
fraction collection method used.
Table 1. XBridge OST C18
Column selection guide for detritylated oligonucleotide purification.
7) How to make select IP-RP buffers (1 liter)
i. Perform all work in a fume hood.
ii. Filter all mobile phases through a solvent compatible, 0.45 µm membrane filter and store in bottles that are clean and particulate free.
In support of customers working on oligonucleotides, Waters offers these critical best practices to ensure consistent high-performance IP-RP liquid chromatographic separations for oligonucleotide analysis. Robust methods are critical to delivering high quality oligonucleotides in diagnostic or therapeutic applications. For example, high quality oligonucleotide primers and probes are essential for accurate PCR-based diagnostic assays, including those for COVID-19.
720006948, Revised December 2020