This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of naphthoic acids in groundwater using Oasis MAX separation method.
Naphthoic acids and naphthalenes in contaminated groundwater is detrmined in this application brief.
Column: |
XTerra MS C18, 2.1 x 100 mm |
Mobile phase A: |
15 mM Ammonium formate |
Mobile phase B: |
ACN |
Flow rate: |
200 μL/min |
Injection volume: |
20 μL |
Instrument: |
Alliance 2695 Separations Module |
Instrument: |
Quattro Ultima |
Ion source: |
Negative Electrospray |
Mode: |
Multiple Reaction Monitoring (MRM) |
Source temp.: |
150 °C |
Desolvation temp.: |
450 °C |
Cone gas: |
50 L/Hr |
Dosolvation gas: |
500 L/Hr |
Collision gas: |
Argon |
1- naphthoic acid: 69% (spike level 0.5μg/L)
2- naphthoic acid: 75% (spike level 0.5μg/L)
Naphthalene: 85% (spike level 100μg/L)
WA31764.105, June 2003