• Application Note

Analysis of Mono-, Di- and Oligosaccharides Using ACQUITY UPLC BEH Amide Columns

Analysis of Mono-, Di- and Oligosaccharides Using ACQUITY UPLC BEH Amide Columns

  • Waters Corporation

This is an Application Brief and does not contain a detailed Experimental section.

Abstract

This application brief describes the analysis of Oligosaccharides using ACQUITY UPLC BEH Amide Columns.

Introduction

Compounds used for this tudy includes:

1.    p-Toluamide

2.   Fructose

3.   Glucose

4.   Sucrose

5.   Maltose

6.   Maltotriose

7.    Maltotetraose

8.   Maltopentaose

9.   Maltohexahose

10.  Maltoheptaose

Experimental

Chromatographic Conditions

Column:

ACQUITY UPLC BEH Amide 2.1 x 50 mm, 1.7 μm

Part Number:

186004800

Mobile Phase A:

80/20 MeCN/H2O with 0.05% triethylamine [TEA]

Mobile Phase B:

30/70 MeCN/H2O with 0.05% triethylamine [TEA]

Flow Rate:

0.17 mL/min

Gradient:

5 minute gradient, 80%-50% MeCN

Injection Volume:

0.7 μL (PLNO)

Sample Concentration:

1 mg/mL each

Sample Diluent:

50/50 MeCN/H2O

Column Temperature:

35 °C

Strong Needle Wash:

20/80 MeCN/H2O (800 μL)

Weak Needle Wash:

75/25 MeCN/H2O (500 μL)

Seal Wash:

50/50 MeCN/H2O

Instrument:

Waters ACQUITY UPLC with ELSD

Gradient:

Time (min)

%A

%B

0.00

100.00

0.00

5.00

40.00

60.00

5.01

100.00

0.00

15.00

100.00

0.00

ELSD Conditions

Gain:

200

Pressure:

40 psi

Drift Tube Temperature:

40 °C

Nebulizer:

Cooling

Data Rate:

10 pps

Filter Time Constant:

Normal

Results and Discussion

WA60110, October 2009