This is an Application Brief and does not contain a detailed Experimental section.
This application brief describes the analysis of Oligosaccharides using ACQUITY UPLC BEH Amide Columns.
Compounds used for this tudy includes:
1. p-Toluamide |
2. Fructose |
3. Glucose |
4. Sucrose |
5. Maltose |
6. Maltotriose |
7. Maltotetraose |
8. Maltopentaose |
9. Maltohexahose |
10. Maltoheptaose |
Column: |
ACQUITY UPLC BEH Amide 2.1 x 50 mm, 1.7 μm |
Part Number: |
186004800 |
Mobile Phase A: |
80/20 MeCN/H2O with 0.05% triethylamine [TEA] |
Mobile Phase B: |
30/70 MeCN/H2O with 0.05% triethylamine [TEA] |
Flow Rate: |
0.17 mL/min |
Gradient: |
5 minute gradient, 80%-50% MeCN |
Injection Volume: |
0.7 μL (PLNO) |
Sample Concentration: |
1 mg/mL each |
Sample Diluent: |
50/50 MeCN/H2O |
Column Temperature: |
35 °C |
Strong Needle Wash: |
20/80 MeCN/H2O (800 μL) |
Weak Needle Wash: |
75/25 MeCN/H2O (500 μL) |
Seal Wash: |
50/50 MeCN/H2O |
Instrument: |
Waters ACQUITY UPLC with ELSD |
Time (min) |
%A |
%B |
---|---|---|
0.00 |
100.00 |
0.00 |
5.00 |
40.00 |
60.00 |
5.01 |
100.00 |
0.00 |
15.00 |
100.00 |
0.00 |
Gain: |
200 |
Pressure: |
40 psi |
Drift Tube Temperature: |
40 °C |
Nebulizer: |
Cooling |
Data Rate: |
10 pps |
Filter Time Constant: |
Normal |
WA60110, October 2009