For research use only. Not for use in diagnostic procedures.
In this application note, we present a novel configuration of the Waters ACQUITY UPLC System with 2D Technology. This minimally modified system is able to utilize the advantages of both HILIC and reversed phase methods in tandem to provide a truly orthogonal and high resolution separation of amphiphilic lipids. We demonstrate the application of 2D UPLC-MS for the analysis of lipids in human plasma as this biofluid is a highly complex matrix with large lipid diversity covering many orders of concentration, thus presenting an analytical challenge when analyzed using traditional single dimensional liquid chromatography.
Lipids play many important roles in maintaining homeostasis of living organisms. These include energy storage, maintaining structural integrity of cell membranes, and acting as signaling molecules. Understanding these lipids may provide insights into mechanisms of disease, including the identification of biomarkers and potential drug targets. Lipids can be either hydrophobic or amphiphilic in nature, with phospholipids being the latter comprised of a hydrophilic phosphate head group and a lipophilic diglyceride tail.
The chemical behavior of amphiphilic lipids has led to the adoption of three main techniques for analysis by liquid chromatography including reversed phase, normal phase, and HILIC separation sciences. While reversed phase chromatography separates these lipids based on their lipophilicity (alkyl chain length and/or degree of saturation), it does not show class distinction especially between classes such as the phosphotidylcholines and sphingomyelins. Normal phase and HILIC chromatography, on the other hand, provide a separation based on the lipid’s head group polarity but provides little separation within the given class.
In this application note, we present a novel configuration of the Waters ACQUITY UPLC System with 2D Technology. This minimally modified system is able to utilize the advantages of both HILIC and reversed phase methods in tandem to provide a truly orthogonal and high resolution separation of amphiphilic lipids. We demonstrate the application of 2D UPLC-MS for the analysis of lipids in human plasma as this biofluid is a highly complex matrix with large lipid diversity covering many orders of concentration, thus presenting an analytical challenge when analyzed using traditional single dimensional liquid chromatography.
System: |
ACQUITY UPLC with 2D Technology |
First dimension column: |
ACQUITY BEH HILIC 2.1 x 100 mm, 1.7 μm |
Second dimension column: |
ACQUITY CSH C18 2.1 x 100 mm, 1.7 μm |
Trap column: |
ACQUITY UPLC BEH C8 VanGuard 130Å 2.1 x 5 mm, 1.7 μm |
Mobile phase A: |
95% ACN 5% H2O 10 mM NH4Ac (pH 5.0) |
Mobile phase B: |
50% ACN 50% H2O 10 mM NH4Ac (pH 5.0) |
UPLC flow rate: |
0.5 mL/min |
Time(min) |
%A |
%B |
---|---|---|
Initial |
100 |
0 |
10.0 |
80 |
20 |
10.1 |
20 |
80 |
13.0 |
20 |
80 |
13.1 |
100 |
0 |
16.0 |
100 |
0 |
Mobile phase A: |
40% ACN 60% H2O 10 mM NH4Ac (pH 5.0) |
Mobile phase B: |
10% ACN 90% IPA 10 mM NH4Ac (pH 5.0) |
UPLC flow rate: |
0.5 mL/min |
Time(min) |
%A |
%B |
---|---|---|
Initial |
100 |
0 |
Fraction Elute (FE)* |
100 |
0 |
FE + 0.10 |
60 |
40 |
FE + 20.0 |
0 |
100 |
FE + 23.0 |
0 |
100 |
FE + 23.1 |
100 |
0 |
FE + 25 |
100 |
0 |
*Fraction Elute (FE) time will vary according to the RT of the lipid class of interest.
First dimension column temp.: |
30 °C |
Second dimension column temp.: |
65 °C |
Time (min) |
Left valve |
Right valve |
---|---|---|
0 |
Position 2 |
Position 2 |
Fraction Trap |
Position 1 |
Position 1 |
Fraction Elute (FE) |
Position 2 |
Position 2 |
FE + 25 |
Position 1 |
Position 1 |
Mass spectrometer: |
Waters Xevo G2 QTof |
Acquisition mode: |
ESI +ve / -ve, MSE |
Capillary voltage: |
2.0 kV (+ve) / 1.00 kV (-ve) |
Sampling cone: |
35.0 V |
Extraction cone: |
4.0 kV |
MS collision energy: |
4.0 V |
MSE energy ramp: |
20 to 45 V |
Source temp.: |
120 °C |
Desolvation temp.: |
500 °C |
Desolvation gas flow: |
1000 L/h |
Cone gas flow: |
10 L/h |
Acquisition range: |
m/z 50 to 1200 |
Lock mass (LeuEnk): |
+ve, m/z 556.2771 and 278.1141 -ve, m/z 554.2615 and 236.1035 |
Lipids were extracted from human plasma using the Waters Ostro sample preparation method.1
By combining the orthogonality of the HILIC and reversed phase separations into a single tandem 2D UPLC-MS method, we overcame the challenges of inter-class co-elution posed by reversed phase liquid chromatography, as shown in Figure 3, between the PC and SM classes. The coelution of these two classes are particularly problematic, as they have only a single dalton difference between them, and co elution makes identification and accurate quantification difficult due to isotopic interferences.
In the HILIC method, inter-class separation is eliminated, as shown in Figure 4, with the PCs and SMs well separated. However, there is now intra-class co-elution which affects the peak capacity and, hence, the sensitivity of the method. In the 2D UPLC-MS method there was improved resolution of the individual lipids of each class using the PC and SM classes as examples, as shown in Figure 5.
The increased peak capacity from the use of combined columns improved both the resolution, shown in Figure 5, as well as the dynamic range of the individual lipids within each class. Using the PC class as an example, the SimLipid (Premier Biosoft) lipid identification software was able to detect and identify 37% more PCs than the HILIC method, and 40% more PCs than the reversed phase C18 method. Generally, the ion intensities for the 2D UPLC-MS method were higher than the HILIC or C18 methods for the same samples run, which could be attributed to the removal of isotopic interferences.
Since the second dimension of separation is reversed phase, the lipids were able to be further separated chromatographically, according to their hydrophobicity (alkyl chain length and degree of saturation), as shown in Figure 6. This was more advantageous than the HILIC method; whereby, residual isotopic interferences due to the co-elution could affect the confidence level for positive identification and quantification of these lipids.
“Heart cutting” a pure fraction of each lipid class using the HILIC column as the first dimension, followed by further separation using reversed phase in the second dimension eliminates errors in identification due to isotopic interferences from co-eluting peaks. This is especially true for the PC and SM classes which differ by 1Da, as shown in Figures 7 and 8.
The limitations posed by traditional single-dimensional separations of HILIC or reversed phase were overcome by the introduction of this novel 2D UPLC-MS configured system which leverages the advantages of both types of methods. This resulted in improved chromatographic resolution, peak capacity, and specificity. In addition, the system is completely automated and UPLC technology provides the capability of high throughput, high resolution analyses compared to traditional HPLCs.
In this application note, we show the orthogonality of this system by pairing the HILIC and C18 chemistries; however, the system can also be easily adapted for other two-dimensional applications such as IEX-RP and SEC-RP. The 2D UPLC-MS system configuration described here uses commercially available components with little modification needed and can easily be switched to other UPLC with 2D technology modes, such as parallel column regeneration or conventional single-dimensional separation.
720004546, June 2013