This application brief demonstrates the UPLC analysis of organic acids.
Column: |
ACQUITY BEH Amide, 2.1 x 100 mm, 1.7 μm |
Part Number: |
186004801 |
Mobile phase A: |
50/50 MeCN/H2O with 10 mM CH3COONH4, pH 9.0 |
Mobile phase B: |
95/5 MeCN/H2O with 10 mM CH3COONH4, pH 9.0 |
Gradient Flow Rate: |
0.6 mL/min |
Injection Volume: |
5.0 μL |
Column Temp: |
50 °C |
Sample Temp: |
5 °C |
Strong/Weak needle wash: |
95/5 MeCN/H2O |
Seal wash: |
10/90 MeOH/H2O |
Instrument: |
ACQUITY UPLC and TQD |
Time (min) |
%A |
%B |
---|---|---|
Initial |
0.1 |
99.9 |
0.4 |
0.1 |
99.9 |
0.5 |
40.0 |
60.0 |
2.0 |
70.0 |
30.0 |
2.01 |
0.1 |
99.9 |
5.0 |
0.1 |
99.9 |
Instrument: |
ACQUITY TQD |
Ionization Mode: |
ES- |
Capillary Voltage: |
4.0 kV |
Cone Voltage: |
-25 V |
Collision Energy: |
10 eV |
Extractor: |
3 V |
RF Lens: |
0.1 V |
Source Temp: |
130 °C |
Desolvation Temp: |
350 °C |
Desolvation Gas: |
650 L/hr |
Cone Gas: |
0 L/hr |
Collision Gas: |
0.1 mL/min |
MRM condition: |
Pyruvic acid: 86.92 > 42.9 Lactic acid: 88.92 > 42.9 Succinic acid: 116.93 > 72.9 Maleic and Fumaric acid: 114.88 > 70.9 |
WA60096, June 2009