This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the UPLC-MS analysis of mono-, di-,and oligosaccharides using ACQUITY UPLC BEH Amide Columns.
Compounds used for this study includes:
1. Fructose
2. Glucose
3. Sucrose
4. Maltose
5. Maltotriose
6. Maltotetraose
7. Maltopentaose
8. Maltohexahose
9. Maltoheptaose
Column: |
ACQUITY UPLC BEH Amide 2.1 x 50 mm, 1.7 μm |
Part number: |
186004800 |
Mobile phase A: |
80/20 MeCN/H2O with 0.10% ammonium hydroxide [NH4OH] |
Mobile phase B: |
30/70 MeCN/H2O with 0.10% ammonium hydroxide [NH4OH] |
Flow rate: |
0.17 mL/min |
Gradient: |
5 minute gradient, 80%–50% MeCN with 10-minute re-equilibration |
Injection volume: |
0.7 μL (PLNO) |
Sample concentration: |
10 μg/mL each |
Sample diluent: |
50/50 MeCN/H2O |
Column temperature: |
35 °C |
Strong needle wash: |
20/80 MeCN/H2O (800 μL) |
Weak needle wash: |
75/25 MeCN/H2O (500 μL) |
Seal wash: |
50/50 MeCN/H2O |
Instrument: |
Waters ACQUITY UPLC with ACQUITY TQD |
Time (min) |
%A |
%B |
---|---|---|
0.00 |
100.00 |
0.00 |
5.00 |
40.00 |
60.00 |
5.01 |
100.00 |
0.00 |
15.00 |
100.00 |
0.00 |
Ionization mode: |
ES |
Capillary: |
2.8 kV |
Cone voltage: |
25 V |
Source temperature: |
120 °C |
Desolvation temperature: |
350 °C |
Desolvation gas flow: |
500 L/Hr |
Cone: |
50 L/Hr |
SIR (m/z): |
179.2 (fructose, glucose); 341.3 (sucrose, maltose); 503.4, 665.5, 827.6, 989.7, 1151.8 (maltooligosaccharides [n=1 to 5]) |
Dwell time: |
0.08 s |
WA60112, Revised August 2021