• Application Note

UPLC-MS/MS Method for Quantitation of EtG and EtS in Human Urine

UPLC-MS/MS Method for Quantitation of EtG and EtS in Human Urine

  • Patrice Y. Ohouo
  • Nicholas Dixon
  • Amy Lieu
  • Zahra Zavery
  • Thomas G. Rosano
  • Clinical and Forensic Toxicology Laboratory, National Toxicology Center
  • Department of Chemistry, University at Albany
  • Department of Basic and Clinical Sciences, Albany College of Pharmacy and Health Sciences
  • Department of Pathology and Laboratory Medicine, Albany Medical Center

For forensic toxicology use only.

Abstract

This application note highlights the development of a rapid, simple dilute and shoot method for the definitive identification and quantitation of ethylglucuronide (EtG) and ethylsulfate (EtS) in human urine using UPLC-MS/MS, for forensic toxicology.

Benefits

Simple dilute and shoot sample preparation method.

Introduction

Ethanol consumption has been linked to significant socio-economic  burdens worldwide.1 As a result, there is a growing need for the detection and identification of ethanol use. Over the years, ethylglucuronide (EtG)  and ethylsulfate (EtS) have emerged as reliable biomarkers of recent ethanol use.2,3 EtG and EtS are minor water soluble phase II metabolites of ethanol and are detectable in urine up to 80 hours following ethanol consumption.2,4 Definitive confirmation of EtG and EtS as a biomarker of ethanol use is performed for a wide range of testing purposes. The authors report the development of a rapid and simple dilute and shoot method for definitive identification and quantitation of EtG and EtS in human urine using UPLC-MS/MS.

Materials

Urine samples 

Human urine samples for the preparation of calibrators and quality  controls (QC) were obtained from volunteer donors with no recent  (at least a week) use of ethanol. Prior to use, samples were confirmed negative for EtG by immunoassay analysis. Authentic samples were collected as part of routine casework. All samples were stored at -20˚C without addition of preservatives.

Reference standards

Drug reference material for EtG (Ethyl-β-D glucuronide, 1.0 mg/mL), and EtS (Ethylsulfate, 1.0 mg/mL) and deuterated  analogues, EtG-D5 (Ethyl-β-D glucuronide D5, 1.0 mg/mL), and EtS-D5 (Ethyl-D5 sulfate, 1.0 mg/mL) were obtained from  Cerilliant Corporation, TX, USA. Deuterated analogues were used for the purpose of internal standardization. Stock solutions containing a mixture of non-deuterated reference material (EtG: 0.1 mg/mL and EtS: 0.05 mg/mL) or a mixture of internal  standard (EtG-D5: 0.1 mg/mL and EtS-D5: 0.05 mg/mL) were prepared in methanol and stored at -20 °C. A daily working  internal standard solution was prepared by a 400-fold dilution of the stock in distilled water.

Experimental

Sample preparation

Urine samples were initially clarified by centrifugation for three minutes at 7200 rpm (~4227 x g). Following centrifugation, 50 µL aliquots of urine were loaded into a 96-well plate (Waters 96-well Sample Collection Plate, 2 mL square well). Aliquots were diluted by adding 500 µL of the daily working internal standard solution. Following dilution, samples were mixed on a vortex for one minute.

LC conditions

LC system:

ACQUITY UPLC I-Class

Column:

ACQUITY UPLC CSH Phenyl-Hexyl 2.1 x 150 mm, 1.7 μm (P/N: 186005408)

Column temp.:

50 °C

Mobile phase A:

Water containing 0.1% formic acid

Mobile phase B:

Acetonitrile

Wash solvent:

Acetonitrile/isopropanol / dH2O (1:1:1) (800 μL)

Purge solvent:

2% methanol in dH2O (2400 μL)

Injection volume:

10 μL

Gradient

Time (min)

Flow rate (mL/min)

%A

%B

Slope

0

0.5

98

2

Initial

0.1

0.5

98

2

6

5

0.5

40

60

6

6.5

0.5

5

95

1

7

0.5

98

2

1

Table 1. Gradient conditions, total run time: 7.5 min.

MS conditions

MS system:

Xevo TQD Mass Spectrometer

Data acquisition and processing:

MassLynx v4.1 with TargetLynx

Ionization mode:

ESI

Capillary voltage:

2.5 kV

Acquisition mode:

Multiple reaction monitoring (MRM – Table 2)

MRM conditions

Compound

Precursor ion (m/z)

Product ion (m/z)

Trace type

EtG

221.1

75.0

Quantifier

EtG

221.1

85.0

Qualifier

EtS

125.0

97.0

Quantifier

EtG-D5

226.1

75.0

Quantifier

EtG-D5

226.1

85.0

Qualifier

EtS-D5

130.0

98.0

Quantifier

Table 2. MRM conditions for EtG, EtS, and corresponding internal standards.

Results and Discussion

A series of calibrators and quality control (QC) samples were prepared by diluting the stock solution of non-deuterated EtG/EtS in negative human urine (Table 3). Following the simple sample preparation, multiple reaction monitoring (MRM) was performed using two transitions for EtG and EtG-D5, and one transition for EtS and EtS-D5 (Figure 1). For EtG a target quantifier/qualifier ion ratio was determined, using the threshold calibrator (EtG/EtS: 500/250 ng/mL), and subsequently used to monitor QC’s and unknown samples. Acceptability criteria included +/- 20% of target ion ratio.

Table 3. Method calibrators and QC’s concentrations and corresponding percent of cut-off (EtG: 500 ng/mL, EtS: 250 ng/mL).
Figure 1. MRM chromatograms from a 10 µL injection of a 500/250 ng/mL EtG/EtS urine calibrator. (A) EtG quantifier ion, (B) EtG qualifier ion, (C) EtG-D5 quantifier ion, (D) EtG-D5 qualifier ion, (E) EtS quantifier ion, (F) EtS-D5 quantifier ion.

Calibration curves were generated based on the ratio of the response of the analyte’s quantifier ion relative to the response of the quantifier ion for the respective deuterated internal standard. Regression lines were plotted using a 1/x weighting. Calibration curves for EtG (r2 range: 0.991–0.999) and EtS (r2 range: 0.997–0.999) were linear over the analytical ranges investigated, and extended from 200 to 10,000 ng/mL and 100 to 5,000 ng/mL for EtG and EtS, respectively (Figure 2). The cut-off for the assay was set at 500 ng/mL for EtG and 250 ng/mL for EtS. The limits of detection (LOD) were determined using the lowest non-zero calibrator approach. LOD’s for EtG and EtS were set at 200 ng/mL and EtS 100 ng/mL, respectively.

The precision and accuracy of the method were assessed at three QC concentrations for EtG (200, 625, 8000 ng/mL) and EtS (100, 312.5, 4000 ng/mL). Based on 11 analytical runs, consisting of three or four replicates, the assay precision (%CV) and accuracy for EtG ranged from 8.4 to 19.6, and 98.4% to 103.6%, respectively. The assay precision and accuracy for EtS ranged from 4.7 to 18.2, and 96.4 to 110.8%, respectively. In all, the method showed good precision and accuracy as summarized in Table 4.

Figure 2. Representative calibration curves for (A) EtG (analytical range: 200 to 10,000 ng/mL) and (B) EtS (analytical range: 100 to 5000 ng/mL).
Table 4. Summary of method precision and accuracy data.

Matrix effects were evaluated using aqueous versus urine based control samples through the analysis of 10 negative urine specimens and aqueous mobile phase spiked with EtG and EtS at 1000 and 500 ng/mL, respectively. Percent matrix effect was calculated using the following formula: [(A/B – 1) × 100%] where A represents the ion response in urine matrix and B represents  the ion response without urine matrix present. Ion effects varied from 1% to -58% for EtG and -54% to 94.6% for EtS. Based on dilute and shoot sample injections, ion suppression of greater than 20% was anticipated, however for this reason analyte-matched deuterated internal standards were incorporated into the method to compensate for matrix effects. Normalization of the data using this approach resulted in a robust assay and satisfied the criteria for precision and accuracy. The stability of EtG and EtS were assessed in both primary specimens and prepared samples following a five day storage period at -10 °C and 4 °C, respectively. Results from reanalysis of primary specimens (n=6), calibrator, and QC samples were within 20% of the results obtained on initial analysis. 

Table 5. Quantitative EtG and EtS results obtained from reference method (MedTox Laboratories, Inc.) and developed method. Data from the developed method was not tabulated (shaded cells) and indicated when quantitation from the reference method was not available (NA).

Method correlation studies were performed using de-identified casework specimens (n=34) with positive presumptive and confirmatory results for EtG and/or EtS. Initial presumptive results were obtained using a qualitative Microgenics DRI® EtG Enzyme immunoassay analysis with a 500 ng/mL cutoff. Quantitative EtG and EtS results were obtained from MedTox Laboratories, Inc. (Minnesota, USA) using a currently validated confirmatory LC-MS/MS method (reference method). Following initial analyses, the specimens were stored at -10 °C for a period of six months. Table 5 shows the results from the reference and developed methods. Analysis of the concentrations obtained by both methods shows dispersion in the data as expected between  laboratories (Figure 3). However, statistical analysis did not show method bias based upon linear regression analysis using 95% confidence limit for the slope and y-intercept (Table 6).

Table 6. Summary of statistical analysis for correlation study between developed and reference methods.
Figure 3. Analysis of the concentrations obtained from developed method compared to reference method (MedTox laboratories, Inc). 

Conclusion

Definitive confirmation of ethanol metabolites is performed as a biomarker of ethanol use for a wide range of testing purposes. The developed method is accurate, precise, and demonstrates the analytical sensitivity necessary for reliable identification of ethanol use based on urine specimen analysis. The simple sample preparation and short chromatography of the developed method can significantly improve the efficiency of EtG and EtS identification and quantitation in a high-throughput setting. 

Acknowledgements

Michelle Wood, Waters Corporation (Wilmslow, UK), Robert Lee, Waters Corporation (Wilmslow, UK), Jonathan Danaceau, Waters Corporation (Milford, MA, USA) and Scott Freeto, Waters Corporation (Beverly, MA, USA) for technical assistance and advice during development of the analytical method.

References

  1. World Health Organization (WHO). Global Status Report on Alcohol and Health. p. XIV. 2014 ed.
  2.  Jaakonmaki PI, Knox KL, Horning EC, Horning MG. Eur. J. Pharmacol. (1967) 1:60–70.
  3. Rosano TG, Lin JJ. Analytical Toxicology. (2008) 32:594–600.
  4. Wurst FM, Skipper GE, Weinmann W. Addiction.(2003) 98 (Suppl 2): 51–61.

720006273, May 2018

Back To Top Back To Top