This paper demonstrates the improvements obtained using sub-2-μm UltraPerformance LC (UPLC) column packing materials UPLC/UV-MS for the analysis of benzenesulfonic acid, p-toluenesulfonic acid, and their alkyl esters.
UPLC combines with PDA and mass detection provides an excellent solution to the analysis of alkyl arylsulfonate esters in drug substances and drug products for genotoxic impurity monitoring. UPLC provides both high resolution and a highthroughput analysis, reducing analysis times from as much as 30 minutes to only 5 minutes. For applications where sensitivity and selectivity are not as demanding, this UPLC/UV system allows for a simple, low cost solution to the analysis of these genotoxic impurities.
Alkyl esters of sulfonic acids, particularly methanesulfonic, benzenesulfonic, and p-toluenesulfonic acid esters, are a common class of reagents used in the pharmaceutical industry as alkylating agents, catalysts, and in purification steps of the chemical synthesis of a drug substance. In addition, sulfonic acids are often used as the final salt form of the drug substance due to improved chemical properties or bioavailability. The presence of any residual alcohols from synthetic reaction or recrystalization steps may result in the formation of alkyl esters of the sulfonic acids. Many of these mesylate, besylate, or tosylate esters are known to be genotoxic, while others are potentially genotoxic, requiring monitoring in the drug substance and drug product.
The U.S. FDA draft guidance1 and the EMEA guidelines2,3 require that any possible genotoxic impurities in a drug substance or drug product that have not been shown to be removed during early synthesis steps be monitored to ensure that the levels are below the Threshold for Toxicological Concern (TTC) of 1.5 μg/day based upon the maximum daily dosage of the pharmaceutical compound.
Depending upon the particular active pharmaceutical ingredient (API), the sensitivity requirements for the analytical method can be quite challenging. For a drug product with a maximum daily dose of 1000 mg, any genotoxic impurity must be analyzed to be less than the TTC level:
Therefore, the genotoxic impurity must be at a level less than 1.5 ppm in the drug product or drug substance.4,5
The most common analytical techniques for monitoring alkyl sulfonate esters have been GC-MS or HPLC/UV-MS with derivatization using pentafluorothiophenol. More recently, HPLC-MS has been shown to give good results without the need for a complicated derivatization step; however, run times on the order of 20 to 30 minutes are required to achieve sufficient resolution from the API.6
This paper demonstrates the improvements obtained using sub-2-μm UltraPerformance LC (UPLC) column packing materials UPLC/UV-MS for the analysis of benzenesulfonic acid, p-toluenesulfonic acid, and their alkyl esters. Analysis times of less than 5 minutes with sufficient sensitivity to meet FDA and EMEA requirements are demonstrated without the need for pre- or post-column derivatization.
LC system: |
Waters ACQUITY UPLC System |
Column: |
ACQUITY UPLC BEH Phenyl Column 2.1 x 50 mm, 1.7 μm |
Column temp.: |
50 °C |
Flow rate: |
600 μL/min |
Mobile phase A: |
5 mM Ammonium Formate, pH 9.0 |
Mobile phase B: |
Methanol |
Gradient: |
10 to 98% B/5 min |
PDA detection: |
210 to 300nm, 1.2 nm resolution, 40 pts/sec |
MS System: |
Waters SQ Mass Detector |
Ionization Mode: |
ESI positive |
Capillary Voltage: |
3200 V |
Cone Voltage: |
24 V |
Desolvation temp.: |
400 °C |
Source temp.: |
130 °C |
Desolvation gas: |
800 L/Hr |
Cone gas flow: |
50 L/Hr |
MS method: |
- ESI+ / SIR |
A method scouting approach was utilized to quickly develop a chromatographic method for the analysis of methyl benzenesulfonate, ethyl benzenesulfonate, methyl p-toluenesulfonate, and ethyl p-toluenesulfonate utilizing an ACQUITY UPLC System with an ACQUITY UPLC PDA detector and a SQ Mass Detector. The MS data was collected in ESI+ mode using timed SIR functions to maximize the dwell time and therefore sensitivity of the analysis. A simple 5-minute gradient from 10% to 98% methanol resulted in a rapid separation with resolution of 1.9 or better for all analytes with sufficient sensitivity to meet FDA and EMEA Genotoxic Impurities Guideline requirements.
A mixture of 1 ppm each of the alkylsulfonate ester standards was analyzed in less than 3 minutes and the UV chromatograms (220 nm) for three replicate injections are shown (Figure 1). All of the alkylsulfonate esters analyzed exhibited good linearity with correlation coefficients greater than 0.9997 for calibration curves covering the concentration range from 0.01 to 10 ppm (Figure 2). This is sufficiently sensitive to monitor these genotoxic and potentially genotoxic impurities (PGIs) in drug substances and drug products at the levels required by FDA and EMEA guidelines as long as the analytes can be well resolved from the active pharmaceutical ingredient and excipients.
Using the conditions described here with 10 μL injections, the retention time reproducibility, and the limits of detection and quantitation determined for standards with UV detection are listed in Table 1. An LOQ of 0.04 ppm corresponds to a quantitation limit of 4 ppm in a 1% solution of a drug substance or a drug product. This is sufficiently sensitive to meet regulatory guidelines for the analysis of these PGIs in pharmaceutical products with a maximum daily dose of 375 mg or less, which is adequate for a number of commercially-available pharmaceuticals. If additional sensitivity is required, preparation of the samples at higher concentrations (~5%) and/or injecting larger volumes (~20 μL) may achieve the desired LOQs, however, the use of MS detection may be necessary to meet the highest sensitivity applications.
*All conc., N=30
Table 1. Limits of detection and quantitation and reproducibility of retention times for methyl and ethyl benzenesulfonates and methyl and ethyl toluenesulfonates standards with this method using UV detection at 220 nm.
Mass spectrometry data (SIR mode) collected for the same 1 ppm standards of the alkylsulfonate esters is shown (Figure 3). The narrow peaks observed with UPLC, which are typically 1 to 3 seconds wide, require rapid MS scanning to obtain 15 to 20 data points across the peaks necessary for accurate and reproducible integration and quantitation. The use of timed SIR functions allows for the optimization of dwell time and sensitivity for each component while maintaining sufficient data points for good integration.
MS data for all of the alkylsulfonate esters analyzed exhibited good linearity with correlation coefficients greater than 0.9960 for calibration curves covering the concentration range from 0.01 to 3 ppm (Figure 4). Using the conditions described here, the retention time reproducibility and the limits of detection and quantitation obtained are listed in Table 2. An LOQ of 6 ppb (MTS) corresponds to a quantitation limit of 0.6 ppm in a 1% solution of a drug substance or a drug product and meets the sensitivity level required for a pharmaceutical product with a maximum daily dose of 2500 mg or less, which is adequate for most commercially available pharmaceuticals.
Higher sample concentrations and larger injection volumes may be employed to achieve additional sensitivity for particularly difficult samples.
The first example shown uses the drug substance Amlodipine Besylate, which is the benzenesulfonic acid salt of Amlodipine. If this material is exposed to residual alcohols from purification/recrystalization steps or any other source, the potential exists for the formation of the potentially genotoxic alcohol esters of the benzenesulfonic acid. This drug substance must therefore be monitored for the presence of these compounds.
Amlodipine Besylate is a long-acting calcium channel blocker use for the treatment of high blood pressure. It is usually administered orally and has a maximum daily dose of 10 mg, which corresponds to a TTC level of 150 ppm for each of the genotoxic impurities. Amlodipine Besylate was obtained from U.S. Pharmacopeia (Rockville, MD) and was analyzed as a 1% solution. Methyl benzenesulfonate (MBS) and ethyl benzenesulfonate (EBS) were not detected in this sample (less than 0.2 ppm) as demonstrated in Figure 5 (MS data).
A second sample was prepared and spiked with methyl benzenesulfonate and ethyl benzenesulfonate at 15 ppm each relative to the solid Amlodipine Besylate representing 1/10th the regulatory requirements for the limits of these genotoxins in this particular drug substance. This sample was analyzed and the resulting MS chromatograms for six replicate injections (Figure 6) demonstrate the excellent sensitivity and reproducibility of the chromatographic method. The separation of the analytes of interest away from the free sulfonic acids ensures that ionization of the sulfonate esters will not be suppressed by the excess of the benzenesulfonic acid. The quantitative MS results (Table 3) display the accuracy and precision of the method with RSDs of 3.5 to 8% and recoveries of 103% and 122% for the spiked MBS and EBS components respectively.
*Average of 6 inj.
Table 3. Analytical MS results for Amlodipine Besylate sample and sample spiked with 15 ppm of methyl benzenesulfonate and ethyl benzenesulfonate.
UV detection has adequate sensitivity to analyze these PGIs in many drug products providing that the TTC levels are sufficiently high and that the analytes of interest are well resolved from the API, the free acid, and any excipients present in the drug product. The analysis using UV detection at 220 nm of the Amlodipine Besylate sample spiked with 15 ppm of MBS and EBS (Figure 7) demonstrates the resolution of the analytes from the benzenesulfonic acid, the late eluting Amlodipine, and other impurities present in the sample. These results illustrate the capability of UV detection to achieve the detection levels required by the FDA and the EMEA for MBS and EBS in this particular sample.
The second example shown uses the drug substance Bretylium Tosylate, which is the p-toluenesulfonic acid salt of Bretylium. Bretylium Tosylate is an antifibrillatory and antiarrhythmic agent that is normally administered by IV and has a maximum daily dose of approximately 2 to 3 gm, which corresponds to a TTC level of 0.5 ppm (for 3 gm dose) for each of the genotoxic impurities. Bretylium Tosylate was obtained from U.S. Pharmacopeia (Rockville, MD) and was analyzed as a 1% solution. Methyl toluenesulfonate (MTS) and ethyl toluenesulfonate (ETS) were not detected in this sample (less than 0.2 ppm) as demonstrated in Figure 8.
A second sample was prepared and spiked with methyl toluenesulfonate and ethyl toluenesulfonate at 0.5 ppm each relative to the solid Bretylium Tosylate representing the regulatory requirements for the limits of these genotoxins in this particular drug substance. This sample was analyzed and the resulting chromatograms for six replicate injections (Figure 9) demonstrate the excellent sensitivity and reproducibility of the chromatographic method. The separation of the analytes of interest away from the free sulfonic acids ensures that ionization of the sulfonate esters will not be suppressed by the excess of the toluenesulfonic acid. The quantitative results (Table 4) display the accuracy and precision of the method with RSDs of 4 to 7% and recoveries of 94% and 104% for the spiked MTS and ETS components respectively.
UV data obtained for this sample (Figure 10) demonstrates the limitations of UV detection for many of these types of genotoxic impurity analyses. Due to the inherent combination of lower sensitivity and specificity of UV detection relative to MS detection, UV detection may be unsuitable for a number of drug products or drug substances with high daily doses that require very high sensitivity or those that are not separated chromatographically from the analytes of interest. Bretylium elutes much earlier in the chromatogram than Amlodipine and tails into the region where the MTS and ETS peaks elute swamping out the analytes.
In addition, the low TTC levels required for this analysis (0.5 ppm in API) are below the quantitation limits of the UV method. Sensitivity could be improved somewhat by injecting a larger volume (20 μL) of a sample concentration of 5%, however, interference of the analytes with the Bretylium peak would still be an issue.
Overall, UPLC with MS detection using SIR mode offers the most sensitive method with the best specificity for the analysis of alkyl arylsulfonates in drug substances and drug products. Despite the limitations of UV detection, UPLC with UV may be used successfully for the analysis of GTIs in a number of pharmaceutical compounds, particularly those with low maximum daily doses that do not require the utmost sensitivity afforded by MS detection.
The ACQUITY UPLC System paired with the ACQUITY UPLC PDA Detector and the SQ Mass Detector provides an excellent system solution to the analysis of alkyl arylsulfonate esters in drug substances and drug products for genotoxic impurity monitoring. UPLC allows for a high resolution and high throughput analysis, reducing analysis times from as much as 30 minutes to only 5 minutes, increasing laboratory productivity while decreasing solvent consumption and the generation of waste solvents.
The use of a single quadrupole mass spectrometer with single ion recording (SIR) methods achieves the specificity and sensitivity necessary to analyze very low levels of impurities in the presence of 1 to 10% concentrations of drug substances or drug products, to meet today’s demanding regulatory requirements for genotoxic impurity analysis. For particular applications where sensitivity and selectivity are not as demanding, UPLC combined with UV detection allows for a simple, low cost solution to the analysis of these genotoxic impurities.