• アプリケーションノート

Separation of Low Levels of Isoleucine from Leucine Using the ACQUITY UPLC H-Class Amino Acid System

Separation of Low Levels of Isoleucine from Leucine Using the ACQUITY UPLC H-Class Amino Acid System

  • Richard C. Daw
  • Waters Corporation

Abstract

In this application note, we describe the chromatographic separation of 17 amino acids in a commercially available amino acid mix within a very short run time, using our ACQUITY UPLC H-Class Amino Acid Analysis System. Baseline resolution of isoleucine and leucine were confirmed a levels as low as 0.05%, meeting regulatory requirements.

Benefits

  • Resolution of leucine and isoleucine at levels as low as 0.05% Ile/Leu using the UPLC Amino Acid Analysis Solution.

Introduction

The European Pharmacacopoeia (Ph. Eur.) defines requirements for the qualitative and quantitative composition of amino acids and mixtures of amino acids. The requirements for allowed impurities are also defined. Manufacturers of amino acids are legally bound to prove that their amino acids meet these specifications before they can distribute their products in Europe.

Leucine (Leu) is a branched-chain α-amino acid and is produced by the fermentation process. During this process, isoleucine can be produced as a by-product. The European Pharmacopoeia states that leucine and isoleucine should have a resolution of 1.5 to levels as low as 0.05%. This application note is intended to demonstrate that the Waters ACQUITY UPLC H-Class Amino Acid System can be used to suitably resolve isoleucine from leucine at these low levels.

The Waters ACQUITY UPLC H-Class Amino Acid System combines UPLC separation technology with AccQ•Tag Ultra derivatization chemistry, providing improved resolution and sensitivity, leading to improved sample characterization, all achieved within a shorter analysis time than conventional methodologies.

The amino acids are derivatized with 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate (AQC) under largely aqueous conditions. The derivatives are then separated utilizing the ACQUITY UPLC H-Class System, enabling analysts to achieve accurate, precise, and robust amino acid analysis utilizing a reversedphase separation, and quantification with either UV or fluorescence detection.

Experimental

Standards, reagents, separation column, and turnkey methodologies within Empower Software projects, are sold as a system solution.

Sample preparation

To a leucine solution, different amounts of isoleucine were spiked to prepare isoleucine/leucine mixtures at 0.0, 0.05, 0.1, and 0.2%. A calibration standard and samples were prepared by transferring 70 μL Borate buffer and 10 μL of the standard/sample to a Waters total recovery vial, vortexing to mix. The derivatization reagent was dissolved in 1 mL of acetonitrile and then 20 μL of the solution was transferred to each vial. Each vial was capped, vortexed, and then heated to 55 °C for 10 minutes prior to analysis.

LC conditions

System:

ACQUITY UPLC H-Class

Detector:

ACQUITY UPLC TUV at 260 nm

Column:

AccQ•Tag Ultra C18, 2.1 x 100 mm, 1.7 μm

Sample temp.:

20 °C

Column temp.:

43 °C

Injection vol.:

0.8 μL

Flow rate:

0.7 mL/min

Mobile phase A:

AccQ•Tag Eluent A

Mobile phase B:

90/10 Water/Eluent B

Mobile phase C:

Water

Mobile phase D:

AccQ•Tag Eluent B

Data management:

Empower 3 Software, SR2

Gradient

Time

 %A

 %B 

%C 

%D 

Curve

0

10

0

90

0

N/A

0.29

9.9

0

90.1

0

11

5.49

9

80

11

0

7

7.1

8

15.6

57.9

18.5

6

7.3

8

15.6

57.9

18.5

6

7.69

7.8

0

70.9

21.3

6

7.99

4

0

36.3

59.7

6

8.59

4

0

36.3

59.7

6

8.68

10

0

90

0

6

10.2

10

0

90

0

6

Results and Discussion

A calibration standard of 17 amino acids was prepared. The standard consisted of a single point calibration curve with each standard at a concentration of 50 pmoles/μL (except cystine at 25 pmoles/μL). As can be seen from the figure below, isoleucine and leucine were resolved at around 7.8 minutes.

The 0.0% Ile/Leu (unspiked) was analyzed and the leucine sample was found to be free of interferences at the retention time of isoleucine. The spiked Ile/Leu samples at 0% (Black line) 0.05% (Blue line), 0.1% (Brown line), and 0.2% (Green line) were analyzed and the peaks were found to have a USP resolution of 2.0 for the 0.05%, 0.1%, and 0.2% levels.

Figure 1. Standard chromatogram.
Figure 2. 0%, 0.05%, 0.1%, and 0.2% Ile/Leu chromatogram overlay.
Figure 3. 0%, 0.05%, 0.1%, and 0.2% Ile/Leu chromatogram overlay (zoomed).

The Ile area was found to be linear when compared to the %Ile/Leu, with an R2 of 0.9999.

Figure 4. Linearity of Ile area vs. %Ile/Leu.

Conclusion

The Waters ACQUITY UPLC H-Class System provided chromatographic separation of all 17 amino acids in a commercially available amino acid mix within a very short run time. Baseline resolution of isoleucine and leucine was confirmed at levels as low as 0.05% Ile/Leu, meeting the regulatory requirements for these components.

720005263, January 2015

トップに戻る トップに戻る