• Nota de aplicación

UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research

UPLC-MS/MS Analysis of 45 Amino Acids Using the Kairos Amino Acid Kit for Biomedical Research

  • Padhraic Rossiter
  • Jaime Salcedo Dominguez
  • Jennifer Warren
  • Norma Breen
  • Lisa J. Calton
  • Waters Corporation

For research use only. Not for use in diagnostic procedures.

Abstract

This study describes the use of the Kairos Amino Acid Kit for analysis of 45 amino acids in solution. Chromatographic separation was performed using an ACQUITY UPLC I-Class System using a CORTECS UPLC C18 Column, followed by detection on a Xevo TQ-S micro System.

Kairos Amino Acid Kit enables biomedical researchers to achieve trusted results through a single flexible kit that allows them to accurately quantitate 42 amino acids within their normal physiological range in less than 10 minutes using the ACQUITY UPLC I-Class/Xevo TQ-S micro System.

Benefits

  • Analytical selectivity of the chromatographic method provides separation of isobaric species
  • Fast analytical run times (<10 minutes)
  • Confidence in peak detection
  • Greater flexibility, system can be used for other analyses

Introduction

The research only Kairos Amino Acid Kit is designed for biomedical researchers to enable the analysis of up to 45 biologically relevant amino acids in less than 10 minutes. The kit calibrators have been value assigned using higher order reference material where available to provide added confidence to your results. Metrological traceability of the Kairos Amino Acid Kit calibrators were traceable to National Metrology Institute of Japan (NMIJ CRM 6011a-6018a, 6022a) for 9 amino acids and further 8 were traceable to National Institute of Standards and Technology (NIST SRM 2389a), the remaining amino acids were gravimetrically prepared using TraceCERT standards, where available, from Sigma-Aldrich.

Here we describe the use of the Kairos Amino Acid Kit for analysis of 45 amino acids in solution. Chromatographic separation was performed using an ACQUITY UPLC I-Class System using a CORTECS UPLC C18 Column, followed by detection on a Xevo TQ-S micro System (Figure 1). Performance of the kit and analytical system was assessed in solution, in addition, NIST SRM 2389a material and matrix samples (urine and plasma) were analyzed for biomedical research purposes.

Figure 1. The Waters ACQUITY UPLC I-Class/Xevo TQ-S micro System.

The chromatographic conditions used allowed for the separation of leucine, methylhistidine, alanine and aminobutyric acid isobaric species.

Experimental

Reagent Kit

Kairos Amino Acid Kit (p/n: 176004379)

LC conditions

System:

ACQUITY UPLC I-Class (FL) with Column Heater (CH-A)

Needle/loop:

10 μL/10 μL

Column:

CORTECS UPLC C18, 1.6 μm, 2.1 x 150 mm

Pre-column:

In-line filter

Mobile phase A:

Water with 0.1% formic acid

Mobile phase B:

Acetonitrile with 0.1% formic acid

Weak wash solvent:

Mobile phase A

Strong wash solvent:

Mobile phase B

Column temp.:

55 °C

Injection volume:

2 μL

Flow rate:

0.5 mL/min

Method conditions

Time (min)

%A

%B

Curve

Initial

99

1

Initial

1

99

1

6

2

87

13

6

5.5

85

15

6

6.5

5

95

6

7.5

5

95

6

7.6

99

1

6

Table 1. Gradient table for the separation of the amino acids.

Run time: 9 minutes

MS conditions

System:

Xevo TQ-S micro

Acquisition mode:

Multiple Reaction Monitoring (MRM) (see Appendix Table A and B for details)

Polarity:

ESI+

Capillary:

2.0 kV

Source temp.:

150 °C

Desolvation temp.:

1000 °C

Cone:

30 V

Data management

MassLynx v4.1 Software with TargetLynx Application Manager

Kit reconstitution

To reconstitute the freeze-dried Kairos Amino Acid Kit, follow the Care and Use Manual (p/n: 720005448EN). Calibrators and QCs were reconstituted using 2 mL of 0.1 M HCl and mixed at room temperature for a minimum of 30 minutes, ensuring all material is fully dissolved.

Internal standard was reconstituted using 2 mL of water and mix at room temperature for 10 minutes, ensuring all material is fully dissolved. The contents of the vial were transferred to a volumetric flask and made up to 10 mL using 10% sulfosalicylic acid supplied in the kit.

Reagent preparation

To prepare the Kairos Amino Acid Kit reagents, Borate Buffer, and AccQ•Tag Ultra “3X” Derivitization Kit reagent, follow the Care and Use Manual. 

Borate buffer preparation: Add 430 µL of 0.5 M NaOH aq. to the 6 mL of borate buffer provided in the derivatization kit (Reagent 1) Expiry: 3 months. Storage: Room temperature.

AccQ•Tag Ultra “3X” reagent: Reconstitute AccQ•Tag Ultra “3X” (Reagent 2B) in 1.5 mL acetonitrile (Reagent 2A). Heat for 10 minutes at 55 °C and vortex. Expiry 5 days. Storage: Room Temperature. AccQ•Tag Ultra should be stored in the desiccator once reconstituted. 

Sample preparation

Step 1 Add 50 μL of sample to 1.5 mL Eppendorf

Step 2 Add 50 μL of 50 μM Internal Standard

Step 3 Vortex mix for 5 seconds

Step 4 Add 50 μL of water

Step 5 Vortex mix for 5 seconds

Step 6 For high concentration samples only – Add 1000 μL of diluent (0.1 M HCl)

Step 7 Centrifuge for 15 minutes at 9000 g

Step 8 Add 70 μL of Borate buffer to maximum recovery vial

Step 9 Add 10 μL of supernatant into Borate buffer and pipette mix

Step 10 Add 20 μL of AccQ•Tag Ultra “3X” reagent

Step 11 Vortex for 5 seconds

Step 12 Allow sample to stand at room temperature for 1 min

Step 13 Heat for 10 minutes at 55 ÅãC

Step 14 Inject 2 μL

For precision and accuracy studies, panel samples were gravimetrically prepared in 0.1 M HCl at 20, 150, 400, and 700 μM for all amino acids and an additional high concentration panel at 2500 μM was prepared for alanine, glycine, isoleucine, leucine, phenylalanine, valine, serine, threonine, tyrosine, and glutamine.

Results and Discussion

No significant system carryover (detector response was ≤20% of Calibrator 1)  was observed from high concentration samples into subsequent blank injections for all amino acids. 

Precision was determined by preparing and analyzing the panel samples in replicates of five over three separate days (n=15), within day and between day precision performance (%CV) were calculated for each panel. Each panel within-day precision was ≤6.9% CV and the between-day precision performance was ≤16.3% CV (≤13.9% CV when argininosuccinic acid was excluded) and the mean within-day and between-day precision performance was less than ≤13.9% CV as shown in Table 2.

Table 2. Mean within- and between-day precision for the analysis of 45 amino acids in solution.

To assess linearity, the Kairos Amino Acid Kit calibrators were analyzed each day over three days. Regression analysis demonstrated a linear fit using 1/x weighting across the concentration range for the 45 amino acids analyzed (Table 3). For  each amino acid, the coefficient of determination (r2) was >0.99 where the back calculated concentrations of the calibrators  were within ±15% of the nominal value, except for the LLOQ (within ±20%). At least 75% of the calibrators fulfilled this criterion, with the exception of the 3 analytes highlighted in red (hydroxyproline, cystathionine, argininosuccinic acid).

Table 3. Linearity of the Kairos Amino Acid Kit in solution. * denotes deviation at LLOQ.

Sensitivity was assessed by the analysis of samples, prepared in 0.1 M HCl over the range 1–20 µM, in replicates of five on one day. The LLOQ for each analyte was determined when the signal-to-noise ratio of the analyte peak was ≥10:1 and the precision performance of the replicates was ≤20% CV. The LOQ for all 45 amino acids analyzed were ≤5 µM with the exception of S-sulfocysteine, argininosuccinic acid (20 µM), and carnosine (15 µM), as shown in Table 4.

Table 4. Sensitivity of the Kairos Amino Acid Kit in solution using the ACQUITY UPLC I-Class and Xevo TQ-S micro System.

Accuracy was assessed through the analysis of the independently prepared panel samples as well as NIST SRM 2389a which was gravimetrically diluted to a concentration of 400 µM prior to analysis. All Panel samples were ≤ ±20.4% bias (range -18.5 to +20.4%) for the concentration ranges tested, with the exception of the three analytes highlighted in red (hydroxyproline, cystathionine, argininosuccinic acid). The diluted NIST SRM 2389a calculated concentrations were ≤ ±10.0% bias (range -10.0 to +3.7%), for the analytes present in the material, all data is summarized in Table 5.

Table 5. Accuracy of the Kairos Amino Acid Kit in solution.

Analysis of a urine and plasma sample allowed for the analysis of amino acids at physiological levels. Figure 2, shows the Total Ion Chromatogram (TIC) for 11 amino acids present in a plasma sample and the chromatogram inset, highlights the chromatographic separation of gamma-aminobutyric acid, beta-aminoisobutyric acid, and alpha-aminobutyric acid from the same plasma sample.

Figure 2. Total Ion Chromatogram (TIC) for 11 amino acids present in a plasma sample and inset a chromatogram to show the separation of gamma-aminobutyric acid, beta-aminoisobutyric acid, and alpha-aminobutyric acid in the same plasma sample.
Figure 3. Chromatogram to show the quantitation of histidine (188.5 µM), alanine (72.4 µM) and phenylalanine (13.7 µM) in a urine sample. Signal to noise ratios (SNR) and peak areas are shown.

Conclusion

Kairos Amino Acid Kit enables biomedical researchers to achieve trusted results through a single flexible kit that allows them to accurately quantitate 42 amino acids within their normal physiological range in less than 10 minutes using the ACQUITY UPLC I-Class/Xevo TQ-S micro System. 

Using AccQ-Tag Ultra “3X” Derivitization Kit to derivatize the samples allows for a simple, robust, and fast reversed-phase UPLC analysis of the amino acids without the need for mobile-phase buffers or ion-pair reagents. This means that you do not need a dedicated system and so the ACQUITY UPLC I-Class/Xevo TQ-S micro System can be used for other analyses. The derivatized samples are stable and in less than 10 minutes, the chromatographic conditions deliver the separation of isobaric amino acids giving you confidence in peak identification.

The method described in the research only Kairos Amino Acid Kit allows for the quantitative analysis of 42 amino acids in solution using an ACQUITY UPLC I-Class/Xevo TQ-S micro System with good precision, linearity, sensitivity, and accuracy. The remaining three analytes; hydroxyproline, cystathionine, and argininosuccinic acid present in the kit are for semi-quantitative analysis only.

720006487, January 2019