This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of paraquat and diquat in tap water.
Paraquat and diquat are quaternary ammonium herbicides used in great quantities
worldwide for defoliation and weed control (see Figure 1). In the United States
alone, paraquat use was 4 million pounds and diquat use was 20,000 pounds in
2002. Because these compounds are potentially toxic to wildlife and to humans, water samples are monitored to assure that residue
levels meet safety standards.
Column: |
Symmetry C18, 3.9 x 150 mm, 5 μm (p/n: WAT046980) |
Mobile phase: |
25 mM Octanesulfonate, pH 3.0/methanol 45:55 |
Flow rate: |
0.75 mL/min |
Injection volume: |
80 μL |
Sample: |
200 mL spiked tap water processed on Oasis MCX cartridge |
Detection: |
UV @ 227 nm |
WA31763.120, June 2003