In this application, we demonstrate a UPLC-MS/MS method using a CORTECS HILIC Column with significantly improved retention and resolution of both analytes compared with other current HILIC methods.
Diquat and paraquat are doubly charged quaternary ammonium herbicides (Figure 1). They have been and still are extensively used worldwide for controlling both crop and aquatic weeds. The US EPA tolerances for diquat are 100 ppb (µg/kg) in potato and 20 ppb in wheat. The US EPA tolerances for paraquat are 500 ppb in potato and 1100 ppb in wheat Diquat and paraquat are ionic substances that are poorly retained on C18 or other reversed phase LC columns. Ion-paring reagents can be employed to enhance reversed-phase retention but this approach usually results in significant ion-suppression if mass-spectrometry is used for detection and quantitation (LC-MS). An alternative methodology, Hydrophilic Interaction Liquid Chromatography (HILIC) has a distinct advantage for LC-MS determination of diquat and paraquat; ionization efficiency is improved because no ion-pairing reagents are required. In this application, we demonstrate a UPLC-MS/MS method using a CORTECS HILIC Column with significantly improved retention and resolution of both analytes compared with other current HILIC methods.
LC system: |
ACQUITY UPLC H-Class |
Column: |
CORTECS HILIC Column 1.6 μm, 2.1 x 100 mm (p/n 186007106) |
(Isocratic): |
50:50 A/B |
Mobile phase A: |
200 mM Ammonium Formate Buffer at pH 3.7 |
Mobile phase B: |
Acetonitrile |
Injection volume: |
20 μL |
Column temp.: |
30 °C |
Wash solvent: |
50:50 Acetonitrile/water |
Purge solvent: |
50:50 Acetonitrile/water |
Flow rate: |
0.5 mL/min |
Sample vials: |
Polypropylene autosampler vials (p/n 186002642) |
MS system: |
ACQUITY TQD Mass Spectrometer |
Ionization mode: |
Positive Electrospray |
Source temp.: |
150 °C |
Desolvation temp.: |
350 °C |
Desolvation gas flow: |
800 L/hr |
Cone gas flow: |
30 L/hr |
Collision gas flow: |
0.20 mL/min |
Data management: |
MassLynx v4.1 |
Note: Polypropylene containers should be used for sample collection and for all sample preparation steps. Polypropylene auto sampler vials (part number 186002642) are recommended for UPLC analysis.
Sample of fresh potato and processed whole wheat flour were prepared using a procedures described in a recently published method.1 This methodology, available on the web,2 is recommended by the EU for screening commodities for paraquat and diquat residues.
Potato Analysis. Samples (10 g) were weighed into 50 mL centrifuge tubes. Fortified samples were prepared by spiking aqueous standards into the pre-weighed sample followed by 30 minute equilibration. 10 mL of extraction solution (50:50 methanol/0.1 M HCl in water) was then added and the samples were shaken by hand for 2 minutes. The samples were then heated at 80 °C for 15 minutes, allowed to cool and then centrifuged at 4000 rpm (rcf 3250x g) for 4 minutes. An aliquot of the supernatant was filtered using a 45 micron PTFE syringe filter. A 400 µL portion of the filtered sample was diluted to 1.0 mL with acetonitrile prior to UPLC-MS analysis.
Flour Analysis. Samples (4 g) were weighed into 50 mL centrifuge tubes. Fortified samples were prepared by spiking aqueous standards into the pre-weighed sample followed by 30 minute equilibration. 10 mL of reagent water was then added to the samples, the samples were vortexed and allowed to equilibrate for 15 minutes. The samples were then extracted and processed in the same manner as the potato samples discussed above.
Table 1 summarizes the MRM transitions and LC-MS parameters used for this study.
Figure 2 shows typical UPLC-MS/MS extracted ion chromatograms obtained from the analysis of a sample of whole wheat flour fortified to 10 ppb of each analyte prior to all sample preparation steps. The ion-chromatogram obtained for potato was similar.
The CORTECS UPLC HILIC separation was shown to be suitable for use with the fast screening method adopted for use in the EU. Prior to UPLC-MS/MS analysis the extracted sample was simply filtered and diluted. The CORTECS UPLC HILIC method is isocratic (50:50 acetonitrile/buffer). Therefore, the sample extract need only be diluted 1:1 with acetonitrile to achieve good chromatographic performance. The limit of quantitation (LOQ) was below 10 ng/g for diquat in potato and wheat flour. The LOQ for paraquat was below 10 ppb for potato and about 10 ppb for wheat flour.
720004851, November 2013