In this application note an LC-MS/MS method is described for the quantification of nitrofuran metabolite residues in commercially available honey.
Apiculturerelies on antibiotics to prevent disease propagating through the densely populated bee colonies. The overuse of antibiotics in honey bee colonies can cause high levels of residues in honey products, which becomes a public health issue. Additionally, bacteria that have developed resistance to the applied antibiotics can pose an increased threat to both human and animal health. Consequently, antibiotics become less effective against bacteria and there will be fewer alternatives available for the successful treatment of infection. Unscrupulous producers search for these alternative antibiotics such as nitrofurans to treat disease.
Nitrofuran antibiotics metabolize rapidly with an in vivo half-life in the order of hours, making parent drug detection ineffective. An LC-MS/MS method is described for the quantification of nitro furan metabolite residues [AOZ, AMOZ, 1-Aminohydantion(AH), Semicarbizide (SC)] in commercially available honey. The metabolite residues were extracted from the honey samples by first dissolving the honey in HCl. The samples were cleaned, derivatized, and then enriched using Oasis HLB solid-phase extraction (SPE) devices. The metabolite residues were resolved chromatographically using a XTerra MS C18 Column. Positive ion electrospray mass spectrometry (MS) was used to quantify and confirm the parent ion [ M+H]+ and fragments for each target analyte.
The honey samples were prepared for analysis using a two step SPE process (Figure 1). The first step provides a simple pass through clean-up to fractionate the analytes from the bulk of the matrix. This dramatically improves the subsequent derivitazation procedure. The second SPE protocol provides additional clean-up as well as providing a sample enrichment factor of 10 to 1 (2 g Honey into 200 mL).
A 2 g sample of honey was diluted with 5 mL of 0.12 M HCl and prepared using the procedu reoutlined in Figure 1. The eluent was quantitatively collected and 300 mL of 50 mM 2-nitro benzaldehyde in DMSO is added for derivitization. The sample was hydrolyzed and derivatized for 18 h at 37 °C. The sample was cooled to room temperature and adjusted to pH 7 by addition of 6 mL of 0.1 M K2HPO4 prior to SPE step 2.
Instrument: |
Alliance 2695 Separations Module |
Column: |
XTerra MS C18, 3.5 mm, 2.1 3 100 mm |
Part number: |
186000404 |
Flow rate: |
0.20 mL/min |
Mobile phase: |
Isocratic 70% 20 mM ammonium formate pH 4.0, 30% ACN |
Injection volume: |
20.0 mL |
Temperature: |
30 °C |
Instrument: |
Waters Micro mass Quattro Micro |
Interface: |
Positive Electrospray (ESI+) |
Optics: |
Capillary - 2.9 kV |
Extractor: |
4 V |
RF Lens: |
0.1 V |
Source Block temperature: |
150 °C |
Desolvation temperature: |
350 °C |
Analyte |
MRM |
Cone (V) |
Collision (V) |
---|---|---|---|
AOZ |
236→134 |
28 |
12 |
AMOZ |
335→291 |
28 |
12 |
SC |
209→192 |
25 |
10 |
AH |
249→178 |
28 |
15 |
The SPE protocol described in this paper provides sample enrichment and cleanup acceptable for the routine determination of nitrofuran metabolites in honey. Results obtained from fortified honey samples indicate that the limit of quantification (LOQ) was below 300 ng/Kg.
Sample derivatization is necessary for chromatographic retention for the small, polar nitrofuran metabolites. Unfortunately, derivatization in the presence of unwanted matrix reduces the reaction efficiency and may increase matrix side-reactions that can interf e re with the LC analysis. The two-step SPE procedure described in this method was optimized to minimize the matrix effects prior to derivatization. This procedure reduced the undesirable matrix interferences while minimizing the amount of derivatization reagent required to achieve a successful reaction.
WA41847, June 2007