• Application Note

The Application of UPLC-MSE for the Analysis of Bile Acids in Biological Fluids

The Application of UPLC-MSE for the Analysis of Bile Acids in Biological Fluids

  • Elizabeth J. Want
  • Muireann Coen
  • Perrine Masson
  • Hector C. Keun
  • Jake T.M. Pearce
  • Michael D. Reily
  • Donald Robertson
  • Cynthia M. Rohde
  • Elaine Holmes
  • John Lindon
  • Robert S. Plumb
  • Stephen McDonald
  • Jeremy K Nicholson
  • Imperial College
  • Waters Corporation

For research use only. Not for use in diagnostic procedures.

Abstract

This application note describes a new, sensitive UPLC-MS approach developed to measure bile acid reproducibly and reliably in biological fluids. Over 30 individual bile acids were separated and detected in a 5-minute window using an ACQUITY UPLC HSS T3 2.1 x 100 mm, 1.8 μm Column coupled to a Q-ToF Premier Mass Spectrometer.

Benefits

  • UPLC-MSE can be used to measure bile acids reproducibly and reliably in biological fluids. 
  • This method analyzes a wide range of endogenous metabolites to provide both a targeted assay and a global metabolite profiling approach in the same analytical run. 
  • Use of this metabolic approach allows for a more comprehensible interpretation of metabolite changes, and it can be easily extended to other sample types and studies.

Introduction

Individual bile acids are endogenous markers of liver cell function and studies of both qualitative and quantitative bile acid changes have been conducted as a result of liver and intestinal diseases. The measurement of serum bile acid concentrations can provide information pertaining to liver damage, as well as hepatic and biliary tract diseases.1-4 However, traditional chromatographic methods have not typically provided sufficient separation in order to differentiate between structurally similar bile acids. Utilization of the Waters UltraPerformance LC (UPLC) Technology high resolution chromatographic system has greatly improved the abilility to separate metabolites from endogenous matrice. UPLC provides superior resolution, sensitivity, and throughput compared with conventional LC approaches. Using UPLC, previously co-eluting metabolites can be separated and matrix effects, such as ESI ion suppression, are minimized. By combining UPLC with oa-TOF mass spectrometry, both high-resolution and exact mass measurements can be achieved, aiding the identification of metabolites. This application note describes a new, sensitive UPLC-MS approach developed to measure bile acid reproducibly and reliably in biological fluids. Over 30 individual bile acids were separated and detected in a 5-minute window using an ACQUITY UPLC HSS T3 2.1 x 100 mm, 1.8 μm Column coupled to a Q-ToF Premier Mass Spectrometer. Bile acids were extracted from serum using methanol and a gradient elution of water and acetonitrile was employed, which also enabled the detection of a wide range of endogenous metabolites, such as lipids. MSE data were acquired using a patented acquisition method that collects precursor and product ion information for virtually every detectable species in a mixture. This allowed for characteristic metabolite fragmentation information to be obtained in a single analytical run, easily distinguishing glycine and taurine bile acid conjugates. This assay was applied to the study of the hepatotoxin galactosamine (gaIN) in rat. Serum bile acid changes were observed after gaIN treatment, including elevated taurine-conjugated bile acids, which correlated to liver damage severity. This UPLC-MS approach to bile acid analysis offers a sensitive and reproducible tool that will be of great value in exploring both markers and mechanisms of hepatotoxicity.

Experimental

UPLC conditions

LC System:

ACQUITY UPLC

Column:

ACQUITY UPLC HSS T3, 2.1 x 100 mm, 1.8 μm

Column temp.:

40 °C

Sample temp.:

4 °C

Mobile phase A:

Water

Mobile phase B:

Acetonitrile

Flow rate:

0.5 mL/min

Injection volume:

5 μl

Data processing:

MarkerLynx XS Application Manager

Gradient

Time (min)

%A

Curve

0.0

0

0

2.0

0

6

12.0

5

6

17.0

5

6

18.0

100

6

22.0

100

6

22.5

5

6

23.0

0

6

26.0

0

6

MS conditions

MS system:

Q-ToF Premier Negative electrospray mode

Scan range:

50 to 1000 Da

Capillary voltage:

2.4 Kv

Source temp.:

120 °C

Desolvation temp.:

350 °C

Cone voltage:

35 V

Desolvation gas flow:

900 L/hr

Collision energy (CE):

Low CE: 5 eV

High CE: ramp of 10 to 70 eV

Animal dosing

Table 1 details the dosing of 40, six-week old male Sprague Dawley rats. The animals were euthanized 24 hrs after galactosamine (gaIN) or vehicle administration. Serum was isolated from blood samples collected at necropsy from the abdominal vena cava and stored at -40 °C pending analysis.

Table 1. Animal and treatment details. Control = Group 1 and treated = Group 2.

Clinical chemistry and histopathology

Clinical Chemistry Analysis. Serum alanine aminotransferase (ALT), aspartate aminotransferase (AST), and total bilirubin levels were analyzed using a Vitros 950 analyzer (Ortho-Clinical Diagnostics, Rochester, NY).

Histological Analysis. Liver samples were fixed in 10% buffered formalin, embedded in paraffin, sectioned, and stained with hematoxylin and eosin. Liver sections were assigned the following histoscores: 0) absence of hepatocellular necrosis, 1) minimal necrosis, 2) mild necrosis, 3) moderate necrosis, and 4) marked necrosis.

Sample preparation

Ice-cold methanol (150 μL) was added to 50 μL serum and vortexed for 30 seconds. The samples were then incubated at -20 °C for 20 mins, centrifuged at 13,000 rpm for 10 min, and the supernatant was removed to a clean tube. The supernatant was then dried down in a vacuum evaporator (Savant), reconstituted in 100 μL water, and transferred into 96-well 350 μL plates.

Results and Discussion

The UPLC-MS assay developed allowed for the reproducible separation and detection of 24 fully identified bile acids, plus 10 tentatively identified bile acids. All bile acids eluted within a 4 min window (6 to 10 mins) with good separation of individual bile acids, as shown in Figure 1. This method offers shorter analysis time than conventional HPLC methods (data not shown), and therefore allows significantly increased sample throughput.

Figure 1. Excellent separation of bile acids using UPLC-MS. 24 bile acids were fully identified, with 10 putative identifications, some of which are illustrated here.

Bile acids ionize strongly using negative mode ESI, producing a prominent precursor [M-H]- ion in the low energy data and informative fragmentation data in the high energy data. This allows conjugates to be easily distinguished by fragment ion analysis. Glycine conjugates give rise to a diagnostic ion at 74 m/z and taurine conjugates at 79.9, 106.0, and 124.0 m/z respectively, as shown in Figure 2. This information was obtained using MSE, facilitating the determination of the different conjugation classes of bile acids in a single run.

Importantly, in addition to bile acids, this assay allows for the detection of a wide range of endogenous metabolites, providing additional, complementary metabolite information. This is of great utility in metabonomics studies, as sample numbers may be large and it is desirable to obtain maximum information in a single analytical run. This provides reductions in sample volume, throughput time, and solvent consumption.

Figure 2. A) *Taurodeoxycholic acid (m/z 498) in (top) the low collision energy run and (below) the elevated collision energy total ion chromatograms. B) (top) The corresponding low collision energy and (below) high collision energy spectra, showing characteristic fragment ions. The application of this UPLC-MSE approach allows for the identification of over 30 bile acids in a single analytical run, reducing sample analysis time and the amount of sample required.

Figure 3 shows representative base peak intensity (BPI) UPLC-MS chromatograms of serum from a control animal and those treated with galactosamine, demonstrating varying degrees of severity of liver damage. Dramatic serum BA changes were visible after galactosamine treatment, as shown in Figure 3, with obvious increases in the taurine-conjugated BAs. These increases correlated with the extent of liver damage (as determined by histoscore) and also the clinical chemistry (AST and ALT).

Figure 3. Selected region of the base peak ion (BPI) chromatograms of animals after galactosamine treatment, graded by histoscore. HS0 denotes no liver damage, whereas animals with a score of HS4 had marked liver damage. Taurine-conjugated bile acids were greatly elevated as liver damage increased; glycine conjugated bile acids showed some elevation whereas conjugated bile acids were unchanged.

Conclusion

UPLC-MSE offers a reliable and reproducible approach for the analysis of bile acids in biological fluids. This was demonstrated with serum samples applied to a toxicity study, and could easily be extended to other sample types and studies. The employment of a 26-min gradient allowed for the analysis of a wide range of endogenous metabolites, providing both a targeted assay and a global metabolite profiling approach in the same analytical run. This approach enhances the information obtained, leading to a more comprehensive interpretation of metabolite changes.

References

  1. Yousef IM, Bouchard G, Tuchweber B, Plaa GL. Mechanism Involved in BA-Induced Cholestasis. In Toxicology of the Liver. 2nd edition. Plaa GL and Hewitt WR, editors. Taylor & Francis, NY. 1988. 347–382.
  2. Ostrow J D. Metabolism of Bile Salts in Cholestasis in Humans. In Hepatic Transport and Bile Secretion: Physiology and Pathophysiology. N. Tavolini and P. D. Berk, editors. Raven Press, NY. 1993. 673–712.
  3. de Caestecker JS, Jazrawi RP, Nisbett JA, Joseph AE, Maxwell JD, Northfield TC. Direct Assessment of the Mechanism for a Raised Serum Bile Acid Level in Chronic Liver Disease. Eur J Gastroenterol Hepatol. 1995 7(10):955-61.
  4. Hofmann, AF. The Continuing Importance of Bile Acids in Liver and Intestinal Disease. Arch Intern Med. 1999 13-27;159(22):2647-58.

720004007, June 2011

Back To Top Back To Top