This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of alprenolol.
The compound analyzed in this study is alprenolol.
Column: |
Xterra MS C18 2.1 x 30 mm, 3.5 μm |
Part number: |
186000398 |
Mobile phase A: |
0.1% HCOOH, pH 2.5 |
Mobile phase B: |
0.1% HCOOH in ACN |
Flow rate: |
0.2 mL/min to MS |
Isocratic mobile phase composition: |
85% A: 15% B |
Injection volume: |
20 μL of 100 μg/mL |
Temperature: |
Ambient |
Detection: |
MS ESI+, SIR 250.19 |
Instrument: |
Alliance 2795 HT, Micromass ZQ |
Micromass ZQ ESI+ |
|
---|---|
Capillary (KV): |
3 |
Cone (V): |
30 |
Extractor: |
3 |
RF lens: |
0.5 |
Source temp.: |
150 |
Desolvation temp.: |
350 |
Cone gas flow (L/Hr): |
60 |
Desolvation gas flow (L/Hr): |
500 |
LM resolution: |
15 |
HM resolution: |
15 |
Ion energy: |
1 |
Multiplier (V): |
650 |
WA20738.008, June 2002