This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of Triton X-100 Using UltraPerformance Convergence Chromatography.
Provides a rapid, high efficiency separation for Triton X-100.
Non-ionic surfactants such as Triton X-100 are used in cosmetics, industrial materials, and many other products. Their composition has to be monitored because the differences in ethoxy chain length affect the viscosity, solubility, polarity, and other characteristics of the mixture.
They are typically analyzed by HPLC, SFC, and GC. Analysis by GC and HPLC can be very time-consuming, and HPLC may require derivatization for non-UV absorbing surfactants. In some cases, baseline separation for oligomers is still not achieved.
System: |
ACQUITY UPC2 |
Detection: |
Photodiode Array (PDA) PDA 3D Channel PDA, 210 to 400 nm; PDA 2D Channel 222 nm at 4.8 nm resolution (compensated 380 to 480 nm) |
Column: |
ACQUITY UPC2 BEH 2.1 x 50 mm, 1.7 μm |
Mobile phase A: |
CO2 |
Mobile phase B: |
Methanol |
Wash solvents: |
70:30 methanol/isopropanol |
Separation mode: |
Gradient starting at 2% B to 35% over 1.25 min, back to 2% B in 5 s |
Flow rate: |
2.0 mL/min |
UPC2 Manager: |
1500 psi |
Column temp.; |
40 °C |
Injection volume: |
1.0 μL |
Run time: |
2 min |
Sample: |
10 mg/mL Triton X-100 in isopropanol |
Software: |
Empower 3 |
UltraPerformance Convergence Chromatography (UPC2) provides a rapid, high efficiency separation for Triton X-100. Excellent resolution for approximately 20 oligomers is achieved in two minutes using lower temperature than in GC or traditional SFC, making UPC2 more amenable to thermally labile compounds. Compared with normal phase HPLC, there is a significant reduction in the consumption of toxic solvents.
720004500, November 2012