ACQUITY™ UPLC™ I-Class/Xevo™ TQ-S micro IVD System: Analytical Performance for an Organic Acid
For in vitro diagnostic use. Not available in all countries.
This is an Application Brief and does not contain a detailed Experimental section.
Abstract
This document describes a test of the analytical performance of the ACQUITY UPLC I-Class/Xevo TQ-S micro IVD System for the analysis of methylmalonic acid (MMA) in serum.
Introduction
The Waters™ ACQUITY UPLC I-Class/Xevo TQ-S micro IVD System enables the quantification of organic compounds in human biological liquid matrices.
Experimental
The ACQUITY UPLC I-Class/Xevo TQ-S micro IVD System was controlled by MassLynx™ IVD (v4.2) and the data processed using the TargetLynx™ Application Manager. Calibrators were prepared by spiking commercially available material in 1% (w/v) bovine serum albumin (BSA) in phosphate buffered saline (PBS), and quality controls (QCs) in both plasma and serum. Samples were processed using the following conditions:
Sample Preparation Conditions
A 100-μL sample and internal standard were added to an Ostro™ protein precipitation and phospholipid removal plate. Samples were precipitated with 1% formic acid in acetonitrile, mixed in-well and then eluted into a collection plate, evaporated, and reconstituted prior to analysis.
LC Conditions
Column: |
ACQUITY UPLC CSH™ C18, 1.7 μm, 2.1 mm × 100 mm, with in-line filter |
Mobile phase A: |
Water with 0.2% formic acid |
Mobile phase B: |
Acetonitrile with 0.2% formic acid |
Flow rate: |
0.45 mL/min |
Gradient: |
1% B over 0.3 minutes, 1–20% B over one minute, 20% over 0.2 minutes, 95% B over 0.5 minutes |
MS Conditions
Resolution: |
MS1 (0.75 FWHM), MS2 (0.75FWHM) |
Acquisition mode: |
MRM |
Polarity: |
ESI- |
Results and Discussion
Performance characteristics of MMA using the Waters ACQUITY UPLC I-Class/Xevo TQ-S micro IVD system is shown in Table 1.
Analytical sensitivity of the system for analyzing extracted MMA serum samples is illustrated in Figure 1.
Conclusion
The Waters ACQUITY UPLC I-Class/Xevo TQ-S micro IVD System has demonstrated the capability to deliver analytically sensitive and selective performance with excellent precision and accuracy for the analysis of methylmalonic acid.
Disclaimer
The analytical performance data presented here is for illustrative purposes only. Waters does not recommend or suggest analysis of the analytes described herein. These data are intended solely to demonstrate the performance capabilities of the system for analytes representative of those commonly analyzed using liquid chromatography and tandem mass spectrometry. Performance in an individual laboratory may differ due to a number of factors, including laboratory methods, materials used, intra-operator technique, and system conditions. This document does not constitute a warranty of merchantability or fitness for any particular purpose, express or implied, including for the testing of the analytes in this analysis.