This is an Application Brief and does not contain a detailed Experimental section.
This application highlights the gradient separation of organophosphonic acids.
The compounds used in this study are:
Columns: |
ACQUITY UPLC BEH Amide, 2.1 x 100 mm, 1.7 μm |
Part Number: |
186004801 |
Mobile Phase A: |
50/50 MeCN/H2O with 10 mM CH3COONH4and 0.04% NH4OH, pH 9.0 |
Mobile phase B: |
95/5 MeCN/H2O with 10 mM CH3COONH4 and 0.04% NH4OH, pH 9.0 |
Flow Rate: |
0.5 mL/min |
Injection Volume: |
5.0 μL (PLNO) |
Sample Concentration: |
2 μg/mL each |
Sample Diluent: |
75/25 MeCN/MeOH |
Column Temperature: |
65 °C |
Weak Needle Wash: |
95/5 MeCN/H2O |
Instrument: |
Waters ACQUITY UPLC with ACQUITY SQD |
Ionization Mode: |
ES- |
Capillary: |
2.5 KV |
Cone: |
30 V (EMPA, IMPA, PMPA); 40 V (CMPA); 35 V (MMPA) |
Source Temperature: |
120 °C |
Desolvation Temperature: |
400 °C |
Desolvation Gas Flow: |
800 L/Hr |
Cone: |
5 L/Hr |
SIR m/z: |
122.9 (EMPA); 136.95 (IMPA); 179.0 (PMPA); 177.0 (CMPA); 150.95 (MMPA) |
Dwell Time: |
0.1 s |
WA60104, July 2009