This application note we demonstrate the efficacy of the Waters UPLC Amino Acid Analysis Solution to resolve 27 amino acids and an internal standard in less than 10 minutes and apply this capability to amino acid analysis of several beer and ale samples.
Beer is a complex matrix consisting of over 100 components. Water, ethanol and carbohydrates are the major constituents of beers and ales. However, there are many minor compounds, some of which are critical for proper taste and quality. One class of compounds, amino acids, is metabolized by yeast during fermentation, leading to the formation of critical flavor components. Therefore, the monitoring of amino acids is essential to demonstrate product consistency and ensure customer satisfaction.
Current HPLC methods for amino acids require run times that exceed 30 minutes, with poor resolution between many amino acids. Here we shall demonstrate the efficacy of the Waters UPLC Amino Acid Analysis Solution to resolve 27 amino acids and an internal standard in less than 10 minutes and apply this capability to amino acid analysis of several beer and ale samples.
System: |
ACQUITY UPLC with Tunable UV Detector |
Method: |
Cell culture |
Column: |
AccQ•Tag Ultra, 2.1 x 100 mm |
Temperature: |
60 °C |
Injection volume: |
1.0 μL |
Detection: |
UV @ 260 nm |
Data: |
Empower software |
A stock 1000 pmol/μL stock mixed amino acid standard was prepared per the cell culture method.1,2,3 An intermediate 100 pmol/μL mixture was prepared by mixing 100 μL of stock with 900 μL water. The working derivatized standard was prepared by adding 10 μL of the 100 pmol/μL mixture to 70 μL borate buffer followed by 20 μL AQC derivatization reagent in a total recovery vial and mixing well. The mixture was heated for 10 minutes at 55 °C, cooled to room temperature then injected. The concentration is 10 pmol/μL for the analytes of interest except Cystein (Cys) which is 5 pmol/μL.
14 samples of beer and ale were purchased commercially. These included domestic regular, light, non-alcoholic, dark beers, and an imported Belgian ale. Approximately 100 mL of each beer was sonnicated to remove carbonation. If the sample appeared excessively cloudy or turbid, it was filtered through a 0.45 micron hydrophilic filter. 200 μL of each beer and ale was mixed thoroughly with 160 μL water and 40 μL of a 1000 pmol/μL Norvaline (Nva - internal standard) solution. The preparation of the internal standard is described in the Cell Culture Method.1,2,3 This resulted in a 1:2 dilution (400 μL total volume) of the beer made 100 pmol /μL in internal standard.
10 μL of this mixture was then mixed with 70 μL of borate buffer and 20 μL of AQC derivitazation reagent and heated as described in the standard preparation section. This working sample mixture, now a 20 fold dilution of the beer, made 10 pmol /μL in internal standard (similar to the working standard) was injected.
Figure 1 is a chromatogram of the cell culture standard. Table 1 is reproducibility data (RSD) for retention time and area for 5 injections of this standard. An overlay of the chromatograms of several of the beers analyzed is found in Figure 2. Table 2 lists the quantitated amounts for the amino acids in the 14 beer samples tested.
The differences in amino acid content, both qualitative and quantitative, for the samples tested are quite evident. Proline (Pro) was found in all samples tested and at a high level, not surprising given the fact that beer yeast cannot ferment proline. On the other hand, Taurine (Tau) and Hydroxy-L-lysine (HyLys) were absent or at a very low level. In general the darker beers had higher amino acid content than light beers.
Also note that in Figure 2 there are many unidentified peaks, possibly amino acids not included in the standard mixture or other compounds that contain an amino group that would react with the AccQ•Fluor reagent. Since the methodology is fully compatible with mass spectrometry detection, it is possible to positively identify these additional compounds, which may also be of critical importance to product consistency.
We have successfully demonstrated application of the Waters UPLC amino acid analysis solution (AAA) to determine the amino acids found in several different beers and ales. This method demonstrated excellent resolution of all sample components with a 10 minute cycle time. Simple sample preparation and analysis times that are approximately three times faster than traditional HPLC methods make the UPLC AAA solution ideal for demonstrating consistency of beer production.