This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates gradient separation of morphine and morphine-3-β-glucuronide on ACQUITY UPLC BEH HILIC Columns.
The compounds used in this study are:
Column: |
ACQUITY UPLC BEH HILIC, 2.1 x 100 mm, 1.7 μm |
Part Number: |
186003461 |
Mobile Phase A: |
10 mM NH4COOH, 0.2% HCOOH in 50:50 ACN:H2O |
Mobile Phase B: |
10 mM NH4COOH, 0.2% HCOOH in 90:10 ACN:H2O |
Flow Rate: |
0.788 mL/min |
Injection Volume: |
2.1 μL |
Sample Concentration: |
125 μg/mL |
Sample Diluent: |
75:25 ACN:MeOH with 0.2% HCOOH |
Temperature: |
30 °C |
Detection: |
UV @ 280 nm |
Sampling Rate: |
20 pts/sec |
Time Constant: |
0.1 |
Instrument: |
Waters ACQUITY UPLC with ACQUITY TUV |
Time (min) |
Profile |
|
---|---|---|
%A |
%B |
|
0.00 |
0.1 |
99.9 |
0.37 |
0.1 |
99.9 |
1.46 |
99.9 |
0.1 |
1.50 |
0.1 |
99.9 |
2.00 |
0.1 |
99.9 |
WA60136, August 2009