This application note describes a clinical research method utilizing Oasis PRiME HLB μElution Plate technology for the analysis of 17-OHP in serum. Using only 100 μL sample volume, this method provides sufficient analytical sensitivity to analyze low physiological levels of 17-OHP.
17-hydroxyprogesterone (17-OHP) is a glucocorticoid steroid hormone derived from both progesterone via 17-hydroxylase and 17-hydroxypregnenolone via 3β-hydroxysteroid dehydrogenase. 17-OHP is converted to 11-deoxycortisol via 21-hydroxylase, followed by conversion to cortisol via 11β-hydroxylase. Measurement of 17-OHP by immunoassay can be prone to analytical interference as a result of cross reactivity of reagent antibodies with structurally related steroid hormones. Analysis of these structurally-similar steroid hormones by liquid chromatography-tandem mass spectrometry (LC-MS/MS) provides three levels of selectivity: sample preparation, liquid chromatography, and detection by Multiple Reaction Monitoring (MRM) mass spectrometry.
It has previously been demonstrated that Oasis PRiME HLB Solid Phase Extraction (SPE) has been suitable for extracting steroid hormones from serum samples.1 This application note describes a clinical research method utilizing Oasis PRiME HLB μElution Plate technology for the extraction of 17-OHP, which was automated using a liquid handler – the Tecan Freedom EVO 100/4. Chromatographic separation was performed on an ACQUITY UPLC I Class System using an ACQUITY UPLC HSS T3 VanGuard pre-column and an ACQUITY UPLC HSS T3 column, followed by detection on a Xevo TQD Mass Spectrometer (Figure 1). In addition, we have evaluated External Quality Assessment (EQA) samples for 17-OHP to evaluate the bias and therefore suitability of the method for analyzing 17-OHP for clinical research.
System: |
ACQUITY UPLC I-Class (FTN) |
Needle: |
30 μL |
Column: |
ACQUITY UPLC HSS T3, 1.8 μm, 2.1 mm x 50 mm, (P/N 186003538) |
Pre-column: |
ACQUITY UPLC HSS T3 VanGuard, 1.8 μm, 2.1 mm x 5 mm (P/N 186003976) |
Mobile phase A: |
Water with 2 mM ammonium acetate + 0.1% formic acid |
Mobile phase B: |
Methanol with 2 mM ammonium acetate + 0.1% formic acid |
Needle wash solvent: |
80% methanol(aq) |
Purge solvent: |
40% methanol(aq) |
Column temp.: |
50 °C |
Injection volume: |
15 μL |
Flow rate: |
0.60 mL/min |
Gradient: |
See Table 1 |
Run time: |
4.7 minutes |
System: |
Xevo TQD |
Resolution: |
MS1 (0.75 FWHM), MS2 (0.75 FWHM) |
Acquisition mode: |
Multiple Reaction Monitoring (MRM) (see Table 2 for details) |
Polarity: |
ESI+ |
Capillary: |
0.4 kV |
Source temp.: |
150 °C |
Desolvation temp.: |
450 °C |
Inter-scan delay: |
0.01 seconds |
Inter-channel delay: |
0.02 seconds |
MassLynx v4.1 Software with TargetLynx Application Manager
17-OHP certified reference solution and its stable labeled internal standard were purchased from Sigma Aldrich (Poole, UK). Calibrators were prepared in a surrogate matrix of MSG4000 stripped human serum purchased from Golden West Biologicals (Temecula, CA). 17-OHP calibrators were prepared over the range of 0.76–303 nmol/L, with quality controls (QCs) at 1.5 nmol/L, 30 nmol/L, and 227 nmol/L.
To convert SI units to conventional mass units divide by 3.028 (nmol/L to ng/mL).
Sample extraction was performed using a liquid handler. 25 μL of 75 nmol/L 17-OHP-13C3, 200 μL methanol, and 550 μL water were added to 100 μL of sample. The samples were mixed after each reagent addition. Samples were centrifuged for five minutes at 4000 g.
An aliquot of each of the pre-treated samples (600 μL) was loaded into individual wells of the Oasis PRiME HLB μElution Plate and slowly pulled through at low vacuum (100 mbar). Consecutive washes with 200 μL of 0.1% (v/v) formic acid in 35% (v/v) methanol(aq) and 200 μL 0.1% (v/v) ammonia in 35% (v/v) methanol(aq) were performed to reduce potential ionic interference. Analytes were eluted using 45 μL of methanol, followed by 55 μL water.
No interferences were observed at the retention time of 17-OHP when eleven structurally-related compounds were examined (11-deoxycortisol, 21 deoxycortisol, 21-hydroxyprogesterone, testosterone, androstenedione, corticosterone, cortisol, DHEA, DHEAS, epitestosterone, and dihydrotestosterone). The chromatographic selectivity of the column is demonstrated through the baseline resolution of 17-OHP and its isobar: 21 hydroxyprogesterone (11-deoxycorticosterone) (Figure 2).
It was determined that in terms of analytical sensitivity, the method would allow for precise quantification (<20% RSD) at 0.76 nmol/L. Signal-to-noise ratio of the lowest calibration standard was >10:1 on ten separate occasions.
Total precision was determined by extracting and quantifying three replicates of tri-level QC material on two occasions per day over five separate days (n=30). Repeatability was assessed by analyzing three replicates at each QC level. Low, mid, and high concentrations were 1.5, 30, and 227 nmol/L, respectively.
The method was shown to be linear over the range of 0.68–333 nmol/L when different ratios of high and low concentration pools of the analytes were combined and analyzed. In addition, calibration lines in spiked serum were linear with coefficient of determinations (r2) >0.994.
Matrix effect investigations were performed using individual donor serum samples (n=6). The matrix factor calculated is shown in Table 4. Normalized matrix factor calculations, based on the analyte-to-internal standard response ratio, produced similar values to peak area matrix factor – both demonstrating minimal ion suppression.
EQA samples were analyzed for 17-OHP (n=50). The data obtained was compared to the mass spectrometry mean for the samples, and Deming regression was performed. The correlation can be seen in Table 5 and shows excellent agreement with the EQA scheme MS mean. Altman-Bland agreement demonstrated a mean bias of -5.0%.
An analytically sensitive and selective clinical research method has been developed for the analysis of 17-OHP in serum.
Using only 100 μL sample volume, this method provides sufficient analytical sensitivity to analyze low physiological levels of 17-OHP. Automation of the analytical method in combination with sample tracking capabilities of the liquid handler improves laboratory workflow and reduces sample handling, minimizing the potential for operator error.
720005691, May 2016