This application brief demonstrates to develop a method to measure Testosterone and Dihydrotestosterone (DHT) from plasma samples.
Direct measurement of androgens is made possible without derivatization using LC-MS with online SPE.
Testosterone is a steroid hormone from the androgen group and is the principal male sex hormone and an anabolic steroid (Figure 1).
In men, approximately 5% of testosterone undergoes 5α-reduction to form the more potent androgen, dihydrotestosterone (DHT) (Figure 1).
Measuring the male steroid hormones and related metabolites, particularly at the low levels found in pediatrics and females, has proven to be a major challenge. Immunoassays for steroids often suffer from poor sensitivity and specificity. While traditional LC-MS assays require derivatization procedures and/or long and complicated sample preparation protocols.
In the current study, a research method for the measurement of testosterone and DHT in plasma was developed. This method allows for the direct measurement of these two androgens at physiological levels in male and female samples by LC-MS without the need for derivitization. Moreover the method makes use of automated sample prep, minimizing sample pretreatment.
LC-MS System Configuration: |
LC System: ACQUITY UPLC Mass Spectrometer: Xevo TQ-S |
Column: |
ACQUITY UPLC BEH C18, 1.7 μm, 2.1 x 50 mm |
Sample Preparation: |
ACQUITY UPLC Online SPE Manager (OSM) |
SPE: |
MassTrak C8 OSM Cartridge |
Sample Pre-Treatment: |
200 μL Plasma 20 μL Internal Standard (d3 Testosterone; 13C3 DHT) 300 μL Zinc Sulfate:MeOH Solution; Centrifuge to pellet protein; remove supernatant After addition of zinc sulfate:MeOH we filled up to 1.0mL prior to centrifugation. 25 uL was then injected (no transfer of supernatant → needle adjusted so stays above pellet). Dilute to 1 ml with Water Inject 25 μL of final sample |
Solvent A: |
10% MeOH+0.05% Formic Acid |
Solvent B: |
MeOH+0.05% Formic Acid |
Flow rate: |
0.4 ml/min |
Time |
%A |
%B |
---|---|---|
0 |
60 |
40 |
3.0 |
0 |
100 |
4.0 |
0 |
100 |
4.1 |
60 |
40 |
Conditioning: |
Step 1: 0.5 ml MeOH @ 4 ml/min Step 2: 2.0 ml Magic Mix* @ 3 ml/min |
Equilibration: |
1.0 ml Water @ 4 ml/min |
Sample loading: |
1.0 ml Water @ 2 ml/min |
Cartridge wash: |
Step 1: 0.5 ml 80% 0.2% NH4OH/20% MeOH @ 2 ml/min Step 2: 0.5 ml 70% 0.2% Formic Acid/30% MeOH @ 2 ml/min |
Elution: |
Elution Time: 0.8 minutes Run Time: 0.8 minutes |
Clamp flush: |
Step 1: 2.0 ml Magic Mix* @ 3 ml/min Step 2: 1.0 ml Water @ 3 ml/min |
* Magic Mix= MeOH/ACN/IPA/Water+0.2% Formic Acid
The wide variation in testosterone and DHT concentrations as well as the inherent difficulties in measuring these molecules, has made it difficult to develop test methods for them.
To measure these molecules successfully by LC-MS requires an effective sample preparation method. Using an online SPE system greatly accelerated the development of an LC-MS method by quickly identifying and utilizing optimized SPE conditions for isolating these androgens from plasma samples.
The focus of the clinical research method described herein was on quickly and efficiently developing an integrated SPE sample preparation method with high sensitivity mass spectrometry to simultaneously measure testosterone and DHT. The method developed uses online SPE to provide a high quality sample that is relatively free from sample matrix effects. By using LC-MS augmented with online SPE, the direct measurement of androgens, even at low levels, is possible.
The clinical research method developed here provides: