• Application Note

Employing UPLC-UV-ToF MSE and the UNIFI Scientific Information System for the Analysis of the Forced Degradation of Glipizide

Employing UPLC-UV-ToF MSE and the UNIFI Scientific Information System for the Analysis of the Forced Degradation of Glipizide

  • Jayne Kirk
  • Russell J. Mortishire-Smith
  • Yun Wang Alelyunas
  • Jeff Robitaille
  • Jim Henriksen
  • Jeff Goshawk
  • Mark D. Wrona
  • Waters Corporation

Abstract

In this application note, glipizide was subjected to accelerated stress conditions to generate degradation products. The resulting impurities were identified, interrogated and visualised using the UNIFI Scientific Information System.

Forced degradation and impurity identification of glipizide was successfully characterized utilizing UPLC with PDA and HRMS detection on the UNIFI Scientific Information System platform. The ability to acquire accurate MSE data within one injection allowed glipizide and its impurities to be rapidly identified and comprehensively confirmed. UNIFI enabled the use of custom calculations, incorporation of UV (and analogue) traces as well as trend plots to easy visualize and interpret their relationship and stabilities.

Benefits

  • Comprehensive impurity analysis using the UNIFI Scientific Information System.
  • Integration of the PDA and MS detectors, combined with trend plots and custom calculations for a complete and comprehensive high resolution impurity analysis solution.
  • Simultaneous collection of accurate mass precursor and fragment ion data without the need for preselection of precursors (MSE).

Introduction

UltraPerformance Liquid Chromatography (UPLC), photo diode array (PDA) detection characterisation of both known and unknown impurities  is a critical step in the pharmaceutical industry.1, 2 There is a need to more rapidly identify and further characterise these impurities. By employing high resolution mass spectrometry (HRMS) techniques we can establish a comprehensive dual detection and characterization strategy. 

A tightly integrated data package for the acquisition, identification, confirmation, interrogation, reporting, and storing information in libraries provides a powerful approach for impurity profiling, and has proven to be exceptionally beneficial to scientists for the characterization of complex samples. 

A comprehensive precursor and product ion dataset acquired by use of the MSE acquisition mode is demonstrated. MSE simultaneously acquires full scan precursor and product ion spectra from a single acquisition without the requirement to preselect precursor ions. This simplifies data acquisition because it does not require advanced knowledge of the analytes.3 Additionally, it also ensures all ionizable impurities can be detected, and in most cases identified, in the first acquisition. In this technology brief, glipizide was subjected to accelerated stress conditions to generate degradation products.4 The resulting impurities were identified, interrogated and visualised using the UNIFI Scientific Information System.

Experimental

Sample description

Acid and temperature degradation: The acid catalyzed/temperature degradation studies were carried out by adding a small aliquot of formic acid to a solution of glipizide. The sample was then left at 80 °C and aliquots taken at 0, 6, 24, 48, 72, and 96 hour time points. The samples were diluted 1:10 with mobile phase prior to injection.

Method conditions

System:

ACQUITY UPLC I-Class (FTN)

Column:

ACQUITY UPLC BEH C18, 2.1 x 150 mm, 1.7 μm

Run time:

11.0 minutes

Vials:

Waters Maximum Recovery

Column temp.:

45 °C

Sample temp.:

8 °C

Injection volume:

0.5 μL

Flow rate:

0.4 mL/min

Mobile phase A:

water + 0.1% formic acid

Mobile phase B:

acetonitrile + 0.1% formic acid

Gradient:

Time

%A

%B

Curve

0.0

95

5

7.0

40

60

6

8.0

0

100

6

9.0

95

5

11

MSE conditions

MS system:

Xevo G2-XS QTof

Ionization mode:

ESI+

Source temp.:

120 °C

Desolvation temp.:

450 °C

Desolvation gas:

800 L/hr

Reference mass:

Leucine enkephalin [M+H]+ m/z 556.27658

Acquisition range:

m/z 50–1200

Scan time:

0.1 sec

Capillary voltage:

1.0 kV

Cone voltage:

25 V

Collision energy:

Function 1: 6 eV

Function 2:

Ramped 25–45 eV

Data management

Pathway profiling – UNIFI v1.8.2

Results and Discussion

In this study, glipizide (shown in Figure 1) was used as a model compound to demonstrate impurity identification using UPLC-MSE and the UNIFI Scientific Information System.5,6  Two main impurities at m/z 379.1074 and m/z 321.1020 were identified and have also been reported in the literature.4 Using accurate m/z values, UNIFI determined and verified the correct molecular formulas C16H18N4O5S and C14H16N4O3S, respectively. 

Figure 1. Glipizide.

Glipizide and the two impurities identified in the samples by mass, isotopic intensity and isotopic mass are tabulated along with a trend plot (Figure 2). The trend plot details the decrease of glipizide followed by the concomitant riseof the impurities. 

Figure 2. Glipizide and the identified impurities with their response plotted over the experimental time course.

As depicted in Figure 3, the spectrum shows the low energy channel containing precursor ion information and the bottom spectrum shows the high energy MSE channel, containing fragment ion information. Product ion assignment is automatically performed during the analysis, annotated directly on the spectrum, greatly simplifying interpretation. In the high energy data, the ion at m/z 321.1019 has been annotated showing the bond cleavage and the charge retaining portion of the molecule. 

Figure 3. Low and high energy (MSE) spectra and UV plot reported. The fragment ions are automatically assigned based on the structure (.mol file) of the API.

A custom calculation was used in order to show the % remaining (or conversion of glipizide) over the course of the experiment (Figure 4). The extracted ion chromatogram (XIC) of glipizide in the 24 hr sample (bottom left panel of Figure 4), as well as the PDA trace at 254 nm and a summed trace of all the identified components, is displayed in the component table representing data from all of the injections. 

Figure 4. Response and percentage remaining of glipizide over the experimental time course.

For further characterization, it is possible using the UV or MS, to determine the relative response of the API and any impurities identified within the samples. Custom calculations were implemented (Figures 5 and 6) displaying the percentage of glipizide against the two impurities within the 48 hr sample, respectively.

Figure 5. Building a custom calculation within the editor for calculating the relative intensity of the API and identified impurities within a sample. This can be based on MS or UV response.
Figure 6. Use of a custom calculation displaying the relative intensity of glipizide and the impurities within a sample.

Conclusion

Forced degradation and impurity identification of glipizide was successfully characterized utilizing UPLC with PDA and HRMS detection on the UNIFI Scientific Information System platform. The ability to acquire accurate MSE data within one injection allowed glipizide and its impurities to be rapidly identified and comprehensively confirmed. UNIFI enabled the use of custom calculations, incorporation of UV (and analogue) traces as well as trend plots to easy visualize and interpret their relationship and stabilities.

References

  1. MD Jones and RS Plumb. The application of sub-2-μm particle liquid chromatography operated high mobile linear velocities coupled to orthogonal accelerated time-of-flight mass spectrometry for the analysis of ranitidine and its impurities. Journal of Separation Science. 29: 2409-2420; 2006.
  2. AL Freed, U Kale, H Ando, DT Rossi, and CA Kingsmill. Improving the detection of degradants and impurities in pharmaceutical drug products by applying mass spectral and chromatographic searching. Journal of Pharmaceutical and Biomedical Analysis. 35 (4): 727–738; 2004.
  3. An overview of the principles of MSE, the engine that drives MS performance. Waters White Paper No. 720004036en. October 2011.
  4. G Bansal, M Singh, KC Jindal, and S Singh. LC and LC-MS study on establishment of degradation pathway of glipizide under forced decomposition conditions. Journal of Chromatographic Science. 46: 510-517; 2008.
  5. A novel approach using UPLC-ToF MSE and the UNIFI Scientific Information System to facilitate impurity profiling of pharmaceuticals. Waters Application Note. 720005604en. March 2016.
  6. Methodology for identification of pesticide metabolites in complex matrices using UPLC-ToF/MSE and the UNIFI scientific information system. Waters Application Note. 720005357en. April 2015.

720005902, January 2017

Back To Top Back To Top